DOI 10.1016/j.jcrs.2004.12.060
Late dislocation of scleral-sutured posterior chamber intraocular lenses.
Аннотация
PURPOSE: To examine a recent series of late scleral-sutured posterior chamber intraocular lens (PCIOL) dislocations to identify possible causes and preventive measures.
SETTING
Price Vision Group, Indianapolis, Indiana, USA.
METHODS
In this retrospective non-comparative interventional case series, 5 consecutive patients received treatment for dislocated scleral-sutured PCIOLs between July 2002 and March 2004. Dislocated lenses were resutured or replaced with another scleral-sutured PCIOL.
RESULTS
Dislocation of scleral-sutured PCIOLs occurred 7 to 14 years after implantation. Four dislocations were spontaneous, and 1 was precipitated by trauma. In each case, the suture affixing 1 or both haptics failed. There was no evidence that the suture had eroded through the tissue or that the knot had untied. Microscopic analysis of an explanted IOL with remnants of the suture attached showed localized degradation and cracking of the polypropylene suture material where it had been embedded in the scleral tissue.
CONCLUSION
Suture-fixated PCIOLs can dislocate due to degradation of the suture material over time. The use of larger diameter (9-0 instead of 10-0) polypropylene suture material and placement of the haptic and sutures in the ciliary sulcus to promote attachment of scar tissue may enhance the long-term stability of scleral-fixated PCIOLs.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.