Usability and microbial validation study of an autonomous UV-C disinfection robot.

de Jager Thomas Roberto, Delzenne Tijn Olivier, Claessen Dennis

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Hospital-acquired infections (HAIs) remain a major challenge in healthcare environments, leading to significant patient morbidity, mortality and increased healthcare costs. Automated ultraviolet C (UV-C) disinfection has been recognized as a potentially effective method to reduce microbial contamination on hard, non-porous, high-touch surfaces, minimizing the errors present in manual disinfection efforts.

AIM: To evaluate the microbiological effectiveness of an autonomous moving UV-C disinfection robot (Omnia UV-C v1.0, Ryberg B.V., Delft, The Netherlands) through microbiological validation and usability assessment.

METHODS

The efficacy of disinfection was assessed by collecting surface swab samples from selected high-touch locations before and after robot disinfection, analysing microbial log reductions of indicator bacteria (Escherichia coli, Bacillus subtilis and Pseudomonas putida). Disinfection doses were measured with analogue dosimeters, and the robot's disinfection reports were analysed to correlate UV-C dose with microbial reduction. Static and dynamic disinfection runs were performed to evaluate the added value of automated movement during disinfection.

RESULTS

The robot achieved microbial log reductions exceeding 4 logs (99.99%) at various locations, meeting established disinfection criteria. Disinfection reports, dosimeter data, and microbial assays showed a strong correlation with dynamic disinfection, providing coverage advantages over static approaches. The disinfection doses measured were consistent with effective microbial decontamination.

DISCUSSION: The autonomous UV-C disinfection robot demonstrated effective microbial decontamination, highlighting its potential to aid in infection control in healthcare settings. Future work will include broader testing for various pathogens and shadowed areas to optimize disinfection coverage.

Для цитирования: de Jager Thomas Roberto, Delzenne Tijn Olivier, Claessen Dennis. Usability and microbial validation study of an autonomous UV-C disinfection robot.. Infection prevention in practice. 2026. DOI: 10.1016/j.infpip.2026.100582.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.