Safety and Performance of INVICTA DF4 Ventricular Defibrillation Lead: Results From APOLLO Prospective Multicenter Study.

Witte Klaus K, Martínez Juan Gabriel, Oliveira Mário M, Barrera Alberto, Izquierdo Maite, Mandurino Cosimo, Trucco Maria Emilce, García Salvador José Juan, Dompnier Antoine, Jean Frédéric, Millán Pablo J Sánchez, Israel Carsten, Martino Maria, Marques Pedro

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Implantable cardioverter-defibrillator (ICD) therapy relies on reliable right ventricular leads for pacing, sensing, and defibrillation. The INVICTA DF4 lead is an active-fixation, quadripolar design intended to optimize implant workflow and long-term safety and performance.

METHODS

APOLLO was a prospective, multicenter, single-arm, observational study enrolling adults across Europe with guideline-based indications for de novo ICD or Cardiac Resynchronization Therapy Defibrillator (CRT-D) implantation who received an INVICTA DF4 lead (single- or dual-coil; 58 or 68 cm). The co-primary endpoints focused on safety and performance at 3 months, while secondary endpoints included long-term electrical variables, arrhythmia termination, and lead-related complications through 24 months.

RESULTS

Between June 4, 2021, and October 19, 2022, 446 patients were implanted across 45 sites in 8 European countries. The INVICTA lead was successfully implanted in 100% of the 446 enrolled patients. Leads were positioned either in the apex in 57.2% or right ventricle (RV) septum in 42.8%. Electrical variables stabilized after discharge and remained stable up to 24 months (M24 mean pacing threshold 0.85 ± 0.33 V at 0.5 ms; sensing 13.39 ± 3.17 mV; pacing impedance 525.5 ± 94.2 Ohms; right ventricular (RV) coil continuity 427.8 ± 82.7 Ohms). Of 86 arrhythmic episodes requiring defibrillation, 97.7% were successfully terminated. Lead-related complication free rate was 97.5% at 30 days, 98.9% at 31-730 days, and 96.4% overall. Mortality was 8.5% (38/446) at two years, of which 63% was due to cardiovascular causes.

CONCLUSIONS

In a large, contemporary European cohort, the INVICTA DF4 ventricular lead demonstrated high implant success, durable electrical performance, effective arrhythmia termination, and low complication rates through 24 months. A trend toward less complication in septal positioning was observed compared to apical placement, with comparable electrical performance.

Для цитирования: Witte Klaus K, Martínez Juan Gabriel, Oliveira Mário M, Barrera Alberto, Izquierdo Maite, Mandurino Cosimo, Trucco Maria Emilce, García Salvador José Juan, Dompnier Antoine, Jean Frédéric, Millán Pablo J Sánchez, Israel Carsten, Martino Maria, Marques Pedro. Safety and Performance of INVICTA DF4 Ventricular Defibrillation Lead: Results From APOLLO Prospective Multicenter Study.. Pacing and clinical electrophysiology : PACE. 2026. DOI: 10.1111/pace.70286.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.