DOI 10.1016/j.ijgo.2010.04.032

Медицинская библиотека · International journal of gynaecology and obstetrics: the official organ of the International Federation of Gynaecology and Obstetrics · 2010

Performance of cryotherapy devices using nitrous oxide and carbon dioxide.

Winkler Jennifer L, Jeronimo Jose, Singleton Jered, Janmohamed Amynah, Santos Carlos

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare temperatures reached by 4 different cryotherapy devices commonly used to treat precancerous cervical lesions in low-resource countries using nitrous oxide (N(2)O) and carbon dioxide (CO(2)) as well as with and without a gas conditioner.

METHODS

Bench testing was conducted using 4 different cryotherapy devices and locally obtained N(2)O and medical- and industrial-grade CO(2) refrigerant with and without a gas conditioner. A thermocouple was used to continuously measure the temperature of the cryotherapy tip. Comparison across the cryotherapy devices was based on the mean and lowest temperatures.

RESULTS

Without the gas conditioner, all of the devices tested reached mean temperatures colder than -50°C with N(2)O, although 2 devices reached warmer temperatures in a proportion of N(2)O tests. Only 2 of the devices reached mean temperatures colder than -50°C with CO(2). One-way analysis of variance identified the device as the dominant factor for the temperature differences, while the gas was not a determinant of temperature variation. The gas conditioner hindered the performance of 2 of the devices, and only improved the performance of 1 device.

CONCLUSION

Both N(2)O and CO(2) reach appropriate freezing temperatures with some cryotherapy devices. Performance of some cryotherapy devices is suboptimal.

Для цитирования: Winkler Jennifer L, Jeronimo Jose, Singleton Jered, Janmohamed Amynah, Santos Carlos. Performance of cryotherapy devices using nitrous oxide and carbon dioxide.. International journal of gynaecology and obstetrics: the official organ of the International Federation of Gynaecology and Obstetrics. 2010. DOI: 10.1016/j.ijgo.2010.04.032.