Case Report: Double-click on the ventricle: ventricular safety pacing triggered by right atrial lead dislodgement in a patient after right atrial tumor resection.

Qin Xiaohan, Fan Jingbo, Chen Taibo, Yang Deyan

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Atrial lead dislodgment is not uncommon after permanent pacemaker implantation, especially in patients with atrial anatomic abnormity. Unique electrocardiography manifestation might give clues for promptly diagnosis. In patients with reconstructed right atrium, optimal pacing strategies and lead stability remain challenging.

CASE PRESENTATION: A 44-year-old woman developed symptomatic sinus node dysfunction after right atrial tumor resection and reconstruction. A dual-chamber pacemaker was implanted, resulting in symptom resolution. At routine follow-up, electrocardiography showed ventricle was intermittently captured by one initial pacing artifact which was immediately followed by another pacing artifact at a fixed interval of 110-ms, consistent with ventricular safety pacing. Device interrogation revealed normal ventricular lead parameters but loss of atrial capture. Imaging confirmed atrial lead dislodgement into the right ventricle. The atrial lead was subsequently extracted and the device was reprogrammed to VVI mode. After 3 months follow-up, the patient remained asymptomatic and was with normal cardiac function.

CONCLUSION

This case highlights the importance of diagnostic role of electrocardiography to detect lead dislodgement. In patents with surgically altered right atrium, individualized pacing strategies, such as single-chamber ventricular pacing, may be appropriate. Novel atrial pacing lead with fix helix or leadless pacing might do some help to reduce the risk of dislodgement.

Для цитирования: Qin Xiaohan, Fan Jingbo, Chen Taibo, Yang Deyan. Case Report: Double-click on the ventricle: ventricular safety pacing triggered by right atrial lead dislodgement in a patient after right atrial tumor resection.. Frontiers in cardiovascular medicine. 2026. DOI: 10.3389/fcvm.2026.1903535.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.