MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ultrasmedbio.2026.07.025

Медицинская библиотека · Ultrasound in medicine & biology · 2026

Characterization of Bubble Cloud Formation and Tissue Ablation Using a High-Frequency Endoscopic Histotripsy System.

Lopez Victor A, Gannon Jessica, Keith Amber, Paul Tamalika, Landry Thomas G, Woodacre Jeffrey K, Brown Jeremy A, Allen Irving C, Vlaisavljevich Eli

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Histotripsy is a non-thermal focused ultrasound (US) ablation strategy that non-invasively destroys tissue by generating a cavitation bubble cloud. Prior work has established the potential of histotripsy for many applications, including the non-invasive treatment of liver tumors. To expand histotripsy beyond the currently available clinical systems for targeting deep abdominal tumors, recent work has developed miniaturized high-frequency devices capable of greater precision using devices within an endoscopic form factor. This work investigates the use of one of these high-frequency histotripsy devices to evaluate cavitation generation and ablation capacity in a variety of ex vivo porcine tissues (brain, liver, kidney, pancreas and lymph node) and in vivo murine liver.

METHODS

A 6.3 MHz histotripsy transducer with an aperture of 10 mm and a focal length of 7 mm was co-registered with a 30 MHz US imaging array. Optical imaging was used to identify the cloud formations for different pulsing cycles, while US imaging was used to monitor ex vivo and in vivo histotripsy ablations in tissue. Histological sections were examined, and statistical analyses were performed on imaging data.

RESULTS

Well-defined bubble clouds ranging from 0.54 ± 0.12 mm to 0.82 ± 0.12 mm (axial) were maintained over the course of 10,000 pulses for 3, 5, 8 and 12 cycles for peak negative pressures (p-) of 33-42 MPa. Larger bubble clouds were generated at a higher p- for all cycles. Pearson correlation coefficients of >0.8 for all cases indicated a strong correlation between US and optical imaging. Histotripsy bubble clouds were identified and maintained on US imaging with hypoechoic zones seen immediately post treatment for all tissue types in both ex vivo and in vivo treatments. Histological results showed complete ablation of the targeted regions in excised porcine brain, kidney, pancreas and lymph nodes. Ex vivo porcine liver exhibited extensive loss of hepatocytes but did retain subtle lobular architecture due to remnants of portal connective tissue. In vivo murine liver showed partial ablation, suggesting that higher treatment doses are required.

CONCLUSIONS

These results demonstrate the potential of miniaturized high-frequency histotripsy devices for precise and effective histotripsy ablation across a wide range of potential applications and establish the potential of this device for use in preclinical murine histotripsy studies.

Для цитирования: Lopez Victor A, Gannon Jessica, Keith Amber, Paul Tamalika, Landry Thomas G, Woodacre Jeffrey K, Brown Jeremy A, Allen Irving C, Vlaisavljevich Eli. Characterization of Bubble Cloud Formation and Tissue Ablation Using a High-Frequency Endoscopic Histotripsy System.. Ultrasound in medicine & biology. 2026. DOI: 10.1016/j.ultrasmedbio.2026.07.025.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление