MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.trsl.2026.10.003

Медицинская библиотека · Translational research : the journal of laboratory and clinical medicine · 2026

Performance Characteristics for Flow Cytometric Enumeration of Peripheral Blood Lymphocyte Subsets under Harmonized Laboratory Conditions: A Multicenter Study.

Xiao Lin, Wang Yuan, Li Yi, Wang Feng, Cheng Liming, Duan Chaohui, Hu Min, Li Mianyang, Liang Zhenlong, Qiao Sai, Qu Chenxue, Shi Bisheng, Wu Weimin, Xu Ying, Ying Binwu, You Ran, Zhang Jun, Wu Xu, Cheng Gong, Wang Hui, Li Min, Cui Wei

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Flow cytometric enumeration of peripheral blood lymphocyte subsets (EPBLS) is essential for disease management, yet standardized validation criteria are lacking. We defined key performance characteristics under harmonized conditions to support criteria development.

METHODS

205 whole blood samples and 54 quality control samples at two levels from nine clinical laboratories across five regions in China were used to measure absolute counts (AC) and relative counts (RC) of lymphocyte subsets using flow cytometric single-platform. Performance characteristics were evaluated based on 2,865 measurements and an integrated statistical approach combining the concordance correlation coefficient, Passing-Bablok regression, and Bland-Altman plots.

RESULTS

Performance characteristics for precision, comparability, and sample stability were established to cover the major clinically relevant lymphocyte subset distributions (RC: 1.32-97.53%; AC: 3.37-6,846.98 cells/μL). Intra-assay and inter-assay coefficients of variation were below 8% and 6% for RC, below 10% and 9% for AC, respectively. A rigorous dual-criteria framework (comparability criteria: consistency and correlation; deviation criteria: absolute maximum bias threshold of <20%), supported by a triple-statistical approach, confirmed agreement across reagent batches, laboratories, and storage time points, with comprehensive bias profiles further reported. Whole blood remained stable for 24 hours post-collection (RC bias <14%; AC bias ≤15%), and processed samples for 48 hours post-staining (RC bias <13%; AC bias <19%).

CONCLUSIONS

This study establishes a comprehensive performance threshold and dual-criteria evaluation system covering precision, comparability, and stability for EPBLS, based on a harmonized multicenter protocol using stratified routine clinical samples, providing a practical reference for laboratory validation and quality specification development.

Для цитирования: Xiao Lin, Wang Yuan, Li Yi, Wang Feng, Cheng Liming, Duan Chaohui, Hu Min, Li Mianyang, Liang Zhenlong, Qiao Sai, Qu Chenxue, Shi Bisheng, Wu Weimin, Xu Ying, Ying Binwu, You Ran, Zhang Jun, Wu Xu, Cheng Gong, Wang Hui, Li Min, Cui Wei. Performance Characteristics for Flow Cytometric Enumeration of Peripheral Blood Lymphocyte Subsets under Harmonized Laboratory Conditions: A Multicenter Study.. Translational research : the journal of laboratory and clinical medicine. 2026. DOI: 10.1016/j.trsl.2026.10.003.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.