Test accuracy of artificial intelligence-based grading of fundus images in diabetic retinopathy screening: A systematic review.
Систематический обзорСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zhelev Zhivko, Peters Jaime, Rogers Morwenna, Allen Michael, Kijauskaite Goda, Seedat Farah, Wilkinson Elizabeth, Hyde Christopher
Аннотация
OBJECTIVES: To systematically review the accuracy of artificial intelligence (AI)-based systems for grading of fundus images in diabetic retinopathy (DR) screening.
METHODS: We searched MEDLINE, EMBASE, the Cochrane Library and the ClinicalTrials.gov from 1st January 2000 to 27th August 2021. Accuracy studies published in English were included if they met the pre-specified inclusion criteria. Selection of studies for inclusion, data extraction and quality assessment were conducted by one author with a second reviewer independently screening and checking 20% of titles. Results were analysed narratively.
RESULTS: Forty-three studies evaluating 15 deep learning (DL) and 4 machine learning (ML) systems were included. Nine systems were evaluated in a single study each. Most studies were judged to be at high or unclear risk of bias in at least one QUADAS-2 domain. Sensitivity for referable DR and higher grades was ≥85% while specificity varied and was <80% for all ML systems and in 6/31 studies evaluating DL systems. Studies reported high accuracy for detection of ungradable images, but the latter were analysed and reported inconsistently. Seven studies reported that AI was more sensitive but less specific than human graders.
CONCLUSIONS: AI-based systems are more sensitive than human graders and could be safe to use in clinical practice but have variable specificity. However, for many systems evidence is limited, at high risk of bias and may not generalise across settings. Therefore, pre-implementation assessment in the target clinical pathway is essential to obtain reliable and applicable accuracy estimates.
METHODS: We searched MEDLINE, EMBASE, the Cochrane Library and the ClinicalTrials.gov from 1st January 2000 to 27th August 2021. Accuracy studies published in English were included if they met the pre-specified inclusion criteria. Selection of studies for inclusion, data extraction and quality assessment were conducted by one author with a second reviewer independently screening and checking 20% of titles. Results were analysed narratively.
RESULTS: Forty-three studies evaluating 15 deep learning (DL) and 4 machine learning (ML) systems were included. Nine systems were evaluated in a single study each. Most studies were judged to be at high or unclear risk of bias in at least one QUADAS-2 domain. Sensitivity for referable DR and higher grades was ≥85% while specificity varied and was <80% for all ML systems and in 6/31 studies evaluating DL systems. Studies reported high accuracy for detection of ungradable images, but the latter were analysed and reported inconsistently. Seven studies reported that AI was more sensitive but less specific than human graders.
CONCLUSIONS: AI-based systems are more sensitive than human graders and could be safe to use in clinical practice but have variable specificity. However, for many systems evidence is limited, at high risk of bias and may not generalise across settings. Therefore, pre-implementation assessment in the target clinical pathway is essential to obtain reliable and applicable accuracy estimates.