Neurosurgery · 2026

Ventriculoperitoneal Shunt Overdrainage Has Transient and Steady-State Components Strongly Dependent on Cerebroventricular Compliance: A Benchtop Experimental Study.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Koueik Joyce, Chebel Roy, Abou Merhi Anas, Pinto Alex, Yang Zhe, Medow Joshua, Luzzio Christopher, Hsu Murielle, Hsu David, Iskandar Bermans J

Аннотация

BACKGROUND AND OBJECTIVES: Using benchtop testing, we previously demonstrated that ventriculoperitoneal shunts are susceptible to chronic overdrainage, not only because of a siphoning effect but also because of intracranial pressure pulsations. These pulsations may be rhythmic (cardiac-related) or sporadic (Valsalva maneuvers) and can drive excess cerebrospinal fluid (CSF) drainage at both horizontal and vertical positions, an effect that is strongly dependent on system compliance. The objective was to characterize the temporal dynamics of CSF flow through a differential pressure valve across extended timescales and physiologically relevant compliance conditions.

METHODS: The pseudo-ventricle benchtop valve testing platform consists of a rigid pseudo-ventricle, a compliance chamber, a pulse generator, and pressure sensors. Flow through a differential pressure shunt valve was measured over 15-minute intervals under controlled simulated physiological conditions, including orientation (horizontal vs vertical), compliance, and pulsation generator force. Each condition was tested in 5 replicates. Data were analyzed using multivariate linear regression, with significance defined as P < .05.

RESULTS: Flow through the valve demonstrated 2 coexisting flow components: (1) a transient exponential component that decays toward zero over time and (2) a steady-state linear component that persists throughout the experiment. The transient component was positive (increased outflow) in the vertical orientation and increased with higher compliance. In the horizontal orientation, it was smaller and could become negative at higher compliance levels. These findings suggest that the transient effect reflects fluid exchange with the compliance chamber during the initial equilibration phase.

CONCLUSION: We identify a third determinant of CSF shunt over drainage, distinct from gravitational siphoning and pulsation-driven pumping. This mechanism involves a posture-dependent transient release of fluid from the compliance chamber on transition to the vertical position, amplified by higher compliance. Effective shunt design must therefore consider not only steady-state siphoning and pulsatile pressures but also this posture-triggered transient compliance response, reinitiated each time a patient moves from horizontal to vertical.