MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1117/1.jbo.31.10.106003

Медицинская библиотека · Journal of biomedical optics · 2026

Phase-aberration-corrected coherence factor beamforming for photoacoustic computed tomography.

Gao Xiali, Zheng Zhichao, Xiao Zhibo, Sun Xu, Zhang Nuomin, Xiao Yang, Huang Hao, Zhou Xuran, Meng Fan, Wang Xuanhao, Chen Ruimin, Shi Junhui

СтатьяПеревод пока не подготовлен

Аннотация

SIGNIFICANCE: Coherence factor (CF) beamforming can suppress sidelobes in photoacoustic computed tomography, but phase aberration can make genuine absorbers appear incoherent and cause conventional CF to downweight real signals.

AIM: We aim to reduce photoacoustic volume tomography phase-aberration-induced loss of true signals in conventional coherence-factor reconstruction and validate the method using phantom and in vivo experiments.

APPROACH: PAC-CF extends an ellipsoidal two-region time-of-flight correction to the angle-weighted channel contributions used for both coherent backprojection and final conventional energy-ratio CF evaluation. The ROI-optimized effective internal sound speed is selected from multiframe PA receive data by maximizing the mean CF within a fixed region of interest.

RESULTS

Simulations showed that conventional CF at the true source dropped to 0.00794 at a 170    m / s sound-speed difference and to ∼ 0.00039 to 0.00042 at larger sound-speed differences, whereas PAC-CF preserved coherence and localization. The four-method phantom ablation showed that combining two-region TOF correction with final CF weighting produced the most concentrated wire responses and strongest background suppression, whereas mouse experiments showed recovery of vascular features suppressed by conventional CF.

CONCLUSIONS

By extending two-region time-of-flight correction to the channel contributions evaluated by the final CF weight, PAC-CF reduces the erroneous suppression of true signals caused by propagation-model mismatch and provides a practical PA-only approach for coherence-weighted three-dimensional PACT.

Для цитирования: Gao Xiali, Zheng Zhichao, Xiao Zhibo, Sun Xu, Zhang Nuomin, Xiao Yang, Huang Hao, Zhou Xuran, Meng Fan, Wang Xuanhao, Chen Ruimin, Shi Junhui. Phase-aberration-corrected coherence factor beamforming for photoacoustic computed tomography.. Journal of biomedical optics. 2026. DOI: 10.1117/1.jbo.31.10.106003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.