Evaluation and correction of charcoal-stripped serum matrix bias in stable-isotope-dilution LC-MS/MS quantification of endogenous serum folates.

Zhu Shuangshuang, Cho Yik Lam, Deng Ying, Wei Meng, Huen Chan Weng, Wong Chui Mae, Yeo Joo Guan, Dong Xueming

СтатьяПеревод пока не подготовлен

Аннотация

Isotope-dilution LC-MS/MS is widely used for targeted quantification of endogenous metabolites, but surrogate-matrix calibration can remain biased when the surrogate matrix is noncommutable with native specimens. In this study, three endogenous serum folate vitamers were quantified by stable-isotope-dilution LC-MS/MS using charcoal-stripped serum (CSS) as a surrogate blank matrix. Post-extraction spiking and pooled-serum standard addition showed analyte-dependent and batch-dependent matrix bias between CSS and native serum. Standard addition comparison revealed matrix factors of 1.005 to 1.611 for 5-methyltetrahydrofolate (5-mTHF), 0.607 to 1.000 for folic acid (FA), and 0.902 to 1.008 for folinic acid, indicating a higher analyte-to-internal-standard normalized response in native serum relative to CSS for 5-mTHF, a lower normalized response for FA, and minimal matrix-dependent response difference for folinic acid. To address this bias, we developed a pooled-serum, batch-specific matrix-factor correction workflow based on standard addition. This approach estimates matrix factors in pooled native serum and applies them to CSS-calibrated concentrations generated under the same batch conditions, avoiding the sample-volume burden of conventional standard addition. Matrix-factor correction reduced validation bias for 5-mTHF and FA to within the predefined ± 15% criterion. In 64 clinical serum samples, correction changed classification relative to selected total folate reference decision limits (5-21 ng/mL) in ten samples (15.6%) for 5-mTHF and five samples (7.8%) for FA. These findings show that CSS-based stable-isotope-dilution LC-MS/MS can produce residual matrix-dependent bias despite normalization using stable-isotope-labeled internal standards (SIL-IS), and that pooled-serum standard addition provides a practical correction strategy.

Для цитирования: Zhu Shuangshuang, Cho Yik Lam, Deng Ying, Wei Meng, Huen Chan Weng, Wong Chui Mae, Yeo Joo Guan, Dong Xueming. Evaluation and correction of charcoal-stripped serum matrix bias in stable-isotope-dilution LC-MS/MS quantification of endogenous serum folates.. Analytical and bioanalytical chemistry. 2026. DOI: 10.1007/s00216-026-06848-w.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.