Feasibility of ultra-low-field MRI in an ion beam treatment room for online MR-guided particle therapy.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Pfitzer Julia, Scharfetter Hermann, Antypas Panagiotis, Georg Dietmar, Fuchs Hermann
Аннотация
BACKGROUND AND PURPOSE: Online MR-guided particle therapy may improve beam delivery precision and treatment efficacy. Ultra-low-field (ULF) MRI may enable image guidance with minimal modification of existing facilities. The feasibility of operating a ULF magnetic resonance (MR) scanner for online MR-guided particle therapy in an unmodified ion-beam therapy environment was investigated.
MATERIAL AND METHODS: A portable 49 mT ULF-MRI system (OSI2 ONE MR scanner) was placed at the isocenter of a research room at an ion-beam therapy facility without radio-frequency shielding. Image quality was assessed using a grid phantom under static and dynamic irradiation conditions for protons and carbon ions. Electromagnetic interference (EMI) was characterized and mitigated via grounding and local shielding. Beam deflections were quantified using a scintillation detector and compared to numerical simulations. The impact of ongoing proton and carbon ion irradiation on the imaging process was assessed. Dosimetric measurements were performed using a Semiflex ionization chamber in a PMMA phantom with and without simultaneous MR imaging.
RESULTS: Stable image quality was achieved across proton and carbon ion irradiation with no relevant irradiation-induced artifacts. Initially substantial EMI was reduced to near-baseline levels by internal shielding. Proton beam deflections ranged from 0.9-1.7 mm at isocenter, consistent with simulations. Dosimetric deviations were below 0.05% for 2D fields and 0.02% for spread-out Bragg peaks (SOBP).
CONCLUSION: ULF-MRI operation in an ion-beam therapy environment without a dedicated shielded room is feasible, without compromising imaging, beam delivery, or dosimetry. This approach supports future MR-guided particle therapy research with reduced infrastructural requirements.
MATERIAL AND METHODS: A portable 49 mT ULF-MRI system (OSI2 ONE MR scanner) was placed at the isocenter of a research room at an ion-beam therapy facility without radio-frequency shielding. Image quality was assessed using a grid phantom under static and dynamic irradiation conditions for protons and carbon ions. Electromagnetic interference (EMI) was characterized and mitigated via grounding and local shielding. Beam deflections were quantified using a scintillation detector and compared to numerical simulations. The impact of ongoing proton and carbon ion irradiation on the imaging process was assessed. Dosimetric measurements were performed using a Semiflex ionization chamber in a PMMA phantom with and without simultaneous MR imaging.
RESULTS: Stable image quality was achieved across proton and carbon ion irradiation with no relevant irradiation-induced artifacts. Initially substantial EMI was reduced to near-baseline levels by internal shielding. Proton beam deflections ranged from 0.9-1.7 mm at isocenter, consistent with simulations. Dosimetric deviations were below 0.05% for 2D fields and 0.02% for spread-out Bragg peaks (SOBP).
CONCLUSION: ULF-MRI operation in an ion-beam therapy environment without a dedicated shielded room is feasible, without compromising imaging, beam delivery, or dosimetry. This approach supports future MR-guided particle therapy research with reduced infrastructural requirements.