Patient Posture-Adaptive Non-Contact Respiratory Rate Measurement Inside a Moving Ambulance Using RGB and Thermal Cameras.

Nemoto Daiga, Takahashi Nozomi, Mukouyama Fumitaka, Nakada Taka-Aki, Nomura Yukihiro, Nakaguchi Toshiya

СтатьяПеревод пока не подготовлен

Аннотация

Background/

Objectives

Monitoring and evaluating respiratory rate (RR) plays a critical role during emergency transport. Conventional RR measurement methods have primarily relied on electrocardiogram-based approaches, which are susceptible to large errors caused by patient body movement and impose a burden on patients due to their contact-based nature. This study aimed to develop a non-contact RR measurement method using RGB and thermal cameras that can be applied inside a moving ambulance.

Methods

One of the key challenges in non-contact RR measurement inside an ambulance is accommodating changes in patient posture. We addressed this by patient posture-adaptive automatic image registration between RGB and thermal cameras and developing a nostril detection method robust to posture changes.

Results

The image registration error of the proposed method was evaluated at various distances from the camera, and the error was confirmed to remain within 1 pixel. RR measurement experiments simulating various respiratory conditions were then conducted inside a moving ambulance. The average of mean absolute error (MAE) was 0.95, 0.59, and 2.80 breaths per minute (breaths/min) for bradypnea, normal, and tachypnea, respectively, in the supine position, and 1.08, 1.90, and 4.38 breaths/min, respectively, in the semi-sitting position. Evaluation of the proposed nostril detection method further confirmed improved RR estimation accuracy compared with the conventional method.

Conclusions

This study achieved patient posture-adaptive automatic image registration and robust nostril detection and successfully conducted experiments under realistic and challenging conditions, representing a substantial step toward the real-world implementation of a non-contact RR measurement system in ambulances.

Для цитирования: Nemoto Daiga, Takahashi Nozomi, Mukouyama Fumitaka, Nakada Taka-Aki, Nomura Yukihiro, Nakaguchi Toshiya. Patient Posture-Adaptive Non-Contact Respiratory Rate Measurement Inside a Moving Ambulance Using RGB and Thermal Cameras.. Diagnostics (Basel, Switzerland). 2026. DOI: 10.3390/diagnostics16193244.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.