Optimization of hemolysis index thresholds for biochemical and immunoassays.
Аннотация
BACKGROUND
Our study aimed at establishing analyte-specific hemolysis-index thresholds on 18 chemistry assays to avoid re-sampling on Atellica© Solution (Siemens Healthcare Diagnostics, Los Angeles, CA, USA).
METHODS
We focused on the analytes which showed the highest degree of hemolysis interference at one or two baseline concentrations for the following 18 analytes: ALT (alanine aminotransferase), ammonium, AST (aspartate aminotransferase), CK (creatine kinase), creatinine, folates, FSH (follicle stimulating hormone), GGT (γ-glutamyltransferase), Hp (haptoglobin), iron, insulin, LD (lactate dehydrogenase), lipase, potassium, urea, uric acid, vitamins B12, and D. Interference was quantified as a percentage difference from the Hb-free baseline sample. Each calculated bias was compared to the four cutoffs: 10%Δ, analytical change limit, total change limit, or reference change limit.
RESULTS
Analytes that did not exceed the predefined allowable-bias criterion for hemolysis included creatinine, FSH, GGT, iron, urea, uric acid, and vitamin B12. Among the affected analytes, the largest positive and negative biases were observed for AST, folates, LD, and potassium. The principal quantitative optimized thresholds for the hemolysis-sensitive analytes were AST, folates, iron, potassium, and LD tests.
CONCLUSION
Our study allowed us to supplement the data provided by the manufacturer and to optimize the thresholds for interference with hemolysis management of hemolyzed samples.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.