Validating a Wearable Bluetooth Oximeter for Infants in the Neonatal Intensive Care Unit.

Courtney Amy, Oei Ju Lee

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Oxygen is vital for life but the wrong amount of oxygen can lead to significant risk of morbidity and mortality. Preterm infants are at particular risk from both too little (hypoxia) and excess (hyperoxia) oxygen. In the neonatal intensive care unit (NICU), oxygen treatment is guided by oximetry derived from wired oximeters which are difficult to use outside of the hospital environment. We aimed to validate Bluetooth wireless oximeters in newborn infants to determine the accuracy of devices that would be more suitable for out of hospital use.

METHODS

Three 20 min sets of measurements including oxygen saturations (SpO2), heart rate (HR) and perfusion index (PI) from the post-ductal regions (left arm and lower limbs) were conducted in 21 spontaneously ventilating infants using the Masimo (1) SafetyNet and (2) Radical 7 oximeters. The wired Radical 7 oximeter was the reference. Data were downloaded onto CSV files and analysed for agreement.

RESULTS

Twenty-one (eight male) infants were recruited between July to September 2023. Median gestations and ages were 36 weeks and 3 days. Neither Bluetooth device (SafetyNet and wireless Radical 7) met International Organization for Standardization (ISO) standard (ARMSE ≤ 3%), Score 3.624 (wireless Rad-7) and 3.884 (SafetyNet). Both wearables consistently overestimated SpO2, with mean biases (wired-wearable) of -1.802 (Rad-7) and -1.803 (SafetyNet), that increased at lower SpO2 levels. Linear regression, ICC, Pearson's correlation analyses showed poor consistency.

CONCLUSION

Wireless oximeters have poor correlation with wired oximeters especially at lower SpO2 levels but ease of use, especially in the home environment in less acutely ill patients, requires further evaluation.

Для цитирования: Courtney Amy, Oei Ju Lee. Validating a Wearable Bluetooth Oximeter for Infants in the Neonatal Intensive Care Unit.. Journal of paediatrics and child health. 2026. DOI: 10.1111/jpc.70622.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.