DOI 10.7759/cureus.116466

Медицинская библиотека · Cureus · 2026

Underestimation of Low-Density Lipoprotein Cholesterol (LDL-C) Using the Friedewald Equation at Higher Triglyceride Levels: Evidence From a Method-Comparison Study.

Gupta Divya, Pakhare Abhijit P, Sindhuja Tulasi, Shukla Rajendra K, Jadhav Ashish A, Chowdhary Rashmi, Kumar Ashok

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

In biochemical laboratories, low-density lipoprotein cholesterol (LDL-C) levels are generally measured using the widely accepted Friedewald equation: [Formula: see text], where TC represents total cholesterol, and HDL stands for high-density lipoprotein. In the case of high serum triglycerides (TGs), the LDL-C value is underestimated using the Friedewald equation. Our study aimed to assess the reliability of calculated LDL-C (cLDL-C) compared with directly measured LDL-C (dLDL-C) in categorizing dyslipidemia patients across different TG levels.

METHODS

cLDL-C was estimated using the Friedewald equation, and dLDL-C was measured by the homogeneous method. Bland-Altman plots (BAPs) were plotted to analyze the agreement between dLDL-C and cLDL-C. We used Lind's concordance correlation coefficient to evaluate the correlation between stroke volume and cardiac output between the two methods. Mean bias with standard deviation and 95% confidence interval was determined at different TG strata.

RESULTS

BAP between dLDL-C and cLDL across TG levels also showed unacceptable agreement at higher TG levels. Mean bias between cLDL-C and dLDL-C is low; however, it rises to +20.2 mg/dL (-17.0 to 57.5) at 250-300 and +26.5 mg/dL (-17.7 to 70.6) above 300 mg/dL. Whereas, beyond TG values >250 mg/dL, cLDL-C shows poor sensitivity. At high TG concentrations, 58.7% of samples at 250-300 mg/dL and 58.1% above 300 mg/dL are assigned to a lower LDL-C band by the Friedewald equation than by direct measurement.

CONCLUSIONS

The Friedewald equation shows acceptable agreement with the homogeneous direct LDL-C assay used in this laboratory at lower TG concentrations and unacceptable agreement at higher concentrations. At high TG concentrations (>250 mg/dL), the Friedewald equation misassigns more than half of patients to a lower LDL-C band than the homogeneous direct measurement and puts them at risk for atherosclerosis. Therefore, we recommend estimating LDL-C using Friedewald equation when serum TG is less than 250 mg/dL; beyond these levels, LDL-C should be measured directly.

Для цитирования: Gupta Divya, Pakhare Abhijit P, Sindhuja Tulasi, Shukla Rajendra K, Jadhav Ashish A, Chowdhary Rashmi, Kumar Ashok. Underestimation of Low-Density Lipoprotein Cholesterol (LDL-C) Using the Friedewald Equation at Higher Triglyceride Levels: Evidence From a Method-Comparison Study.. Cureus. 2026. DOI: 10.7759/cureus.116466.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.