Between-Session Reliability and Agreement of Body-Plethysmographic Lung Volumes in a Legacy Pulmonary Function Cohort.

Izzo Carmine, Izzo Alessandro, Mone Pasquale, Musto Simona, Ciccarelli Michele, Giallauria Francesco

СтатьяПеревод пока не подготовлен

Аннотация

Background

Serial interpretation of lung volumes requires assessing both reliability and agreement. We examined these properties in a historical clinical dataset.

Methods

This retrospective secondary analysis included 72 participants, excluding seven same-session pre/post-bronchodilator pairs. The primary analysis used 19 separately dated pairs labelled pre-bronchodilator at both examinations; the expanded 65-pair analysis was exploratory. Documented intervals were two to three days; the investigator retrospectively confirmed separate-day examinations within six days for all 65 pairs. Total lung capacity (TLC) and residual volume (RV) were principal outcomes; functional residual capacity was secondary and available in the same 19 participants. Analyses included two-way mixed-effects, absolute-agreement, single-measure intraclass correlation coefficients (ICCs), Lin's concordance correlation coefficient, Gaussian reference limits, empirical difference quantiles, and leave-one-out analyses.

Results

Fourteen primary participants had recorded chronic obstructive pulmonary disease labels without independently verified diagnostic post-bronchodilator spirometry. Primary TLC and RV ICCs were 0.947 (95% bootstrap confidence interval 0.838-0.988) and 0.940 (0.818-0.974), respectively. Mean second-minus-first differences were -0.033 and -0.015 L, with Gaussian reference limits of -0.964 to 0.898 L and -0.934 to 0.904 L, respectively. Corresponding empirical 2.5th-97.5th percentiles were -1.122 to 0.641 L and -0.875 to 0.803 L. Removing one observation reduced the standard deviation of paired differences by up to 28.8% for TLC and 14.8% for RV.

Conclusions

High reliability estimates coexisted with appreciable paired variation. Unverified technical quality and clinical stability preclude attributing this variation solely to measurement error or establishing clinical response thresholds.

Для цитирования: Izzo Carmine, Izzo Alessandro, Mone Pasquale, Musto Simona, Ciccarelli Michele, Giallauria Francesco. Between-Session Reliability and Agreement of Body-Plethysmographic Lung Volumes in a Legacy Pulmonary Function Cohort.. Journal of clinical medicine. 2026. DOI: 10.3390/jcm15197704.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.