Prehospital Pediatric Ventilation Quality Following Use of a Feedback Device: A Randomized Study.

Finney Joseph D, Siegler Jeffrey, Wang Jinli, John Kavya, Robinson Debra, Grawey Thomas, McClain Brad, Hoon Lee Sang, Riney Lauren C, Babcock Lynn, Browne Lorin R, Ahmad Fahd A

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Manual ventilation is a lifesaving intervention for children in respiratory arrest, yet the quality of pediatric ventilation delivered by emergency medical services (EMS) clinicians across geographically distinct sites is poorly characterized. We evaluated baseline EMS ventilation performance and whether a ventilation feedback device (VFD) improves adherence to national and international ventilation guidelines for ventilation rate, tidal volume, and mask seal.

METHODS

We conducted a single-anonymized, randomized, crossover study of EMS clinicians from 5 EMS agencies across 3 metropolitan sites in the United States. Using a simulated 3-year-old toddler manikin, participants completed 2 sequential 2-minute respiratory arrest scenarios separated by a 10-minute washout period. Participants performed ventilation with and without real-time VFD feedback in randomized order. The EOLifeX VFD recorded ventilation rate, tidal volume, and mask leak with each breath. Adherence to age-based guidelines was analyzed using generalized linear mixed-effects models with participants nested within sites.

RESULTS

We enrolled 123 participants across 5 agencies. With VFD feedback, guideline-recommended tidal volumes were achieved in 87% of breaths compared with 16.2% without feedback. Ventilation rate adherence improved from 65% to 88.6%. Mask leak of less than 30% occurred in 88.4% of ventilations with feedback vs 80% without feedback.

CONCLUSION

In our simulation study, EMS clinicians rarely met manual ventilation guidelines for pediatric patients in respiratory arrest. The use of a VFD substantially improved guideline adherence, suggesting broad potential to improve ventilation performance.

Для цитирования: Finney Joseph D, Siegler Jeffrey, Wang Jinli, John Kavya, Robinson Debra, Grawey Thomas, McClain Brad, Hoon Lee Sang, Riney Lauren C, Babcock Lynn, Browne Lorin R, Ahmad Fahd A. Prehospital Pediatric Ventilation Quality Following Use of a Feedback Device: A Randomized Study.. Pediatrics. 2026. DOI: 10.1542/peds.2026-077291.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.