DOI 10.1167/tvst.15.10.22

Медицинская библиотека · Translational vision science & technology · 2026

Comparison of Biometric Measurements Between Anterior Segment Hyperparallel and Swept-Source Optical Coherence Tomography.

Servillo Andrea, Porporato Natalia, Tan Bingyao, Xu Benjamin Y, Tun Tin A, Perera Shamira, Husain Rahat, Baskaran Mani, Sultana Rehena, Aung Tin, Schmetterer Leopold

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the inter-device agreement between hyperparallel optical coherence tomography (HP-OCT) and swept-source OCT (SS-OCT) in quantitative anterior segment parameters.

METHODS

Phakic subjects aged ≥40 years were consecutively recruited at a tertiary eye hospital. Horizontal B-scans (0°-180°) were analyzed. On SS-OCT, parameters were measured automatically; on HP-OCT, they were extracted manually using 3D Slicer. Agreement was evaluated using intraclass correlation coefficients (ICCs), Bland-Altman plots, and orthogonal regression.

RESULTS

A total of 118 sectors from 59 subjects were analyzed. ICCs were ≥0.88 for all parameters except iris thickness (IT) at 750 µm (IT750, ICC = 0.66). HP-OCT consistently measured lower values than SS-OCT for angle opening distance (AOD) at 500 µm (AOD500: -0.024 [-0.036 to -0.013]) and 750 µm (AOD750: -0.032 [-0.050 to -0.015]), trabecular-iris space area (TISA) at 500 µm (TISA500: -0.008 [-0.013 to -0.003]) and 750 µm (TISA750: -0.012 [-0.020 to -0.004]), iris curvature (IC: -0.015 [-0.021 to -0.009]), IT750 (-0.031 [-0.041 to -0.022]), and anterior chamber depth (-0.118 [-0.138 to -0.098]). For the following parameters, differences between devices increased with larger measurements, with HP-OCT yielding proportionally lower values (proportional bias): AOD500, AOD750, TISA500, TISA750, IC, and lens vault.

CONCLUSIONS

For anterior segment biometry, HP-OCT and SS-OCT showed high ICCs but limited interchangeability because of systematic and proportional bias.

TRANSLATIONAL RELEVANCE: Understanding how HP-OCT measurements compare with those from SS-OCT is essential for clinical integration, because inter-device agreement and interchangeability guide how HP-OCT can be applied and interpreted in practice.

Для цитирования: Servillo Andrea, Porporato Natalia, Tan Bingyao, Xu Benjamin Y, Tun Tin A, Perera Shamira, Husain Rahat, Baskaran Mani, Sultana Rehena, Aung Tin, Schmetterer Leopold. Comparison of Biometric Measurements Between Anterior Segment Hyperparallel and Swept-Source Optical Coherence Tomography.. Translational vision science & technology. 2026. DOI: 10.1167/tvst.15.10.22.