Fixed-distance pH probe placement produces systematic bias in reflux detection: the role of esophageal length.
Аннотация
PURPOSE: Ambulatory esophageal pH monitoring is the reference standard for diagnosing gastroesophageal reflux disease, yet a subset of symptomatic patients demonstrate physiologically nonpositive studies. To evaluate whether variation in esophageal length and relative probe position influences measured reflux exposure and contributes to symptom-physiology discordance during wireless pH monitoring.
METHODS
We performed a retrospective analysis of data from 63 patients who underwent Bravo pH monitoring system (Medtronic). Relative probe position was defined as the ratio of capsule placement distance to esophageal length. Associations between anatomic variables and acid exposure time (AET) were assessed using multivariable regression, including analyzes of overall, upright, and supine reflux exposure.
RESULTS
Sixty-three patients were included (60% female). Height was strongly associated with esophageal length (β = 0.17; P <.001), resulting in systematic differences in relative probe positions. A subset of patients with an elevated symptom burden demonstrated physiologically nonpositive reflux testing; these patients were more often female and shorter in stature. In multivariable models, taller height was independently associated with increased AET (β = 0.14; P =.02), an effect observed in upright reflux (β = 0.18; P =.030) but not in supine reflux. Apparent sex differences in reflux detection were explained by underlying anatomic variation rather than sex itself.
CONCLUSION
Fixed-distance pH probe placement results in variable relative positioning within the esophagus, producing nonequivalent physiological sampling across patients. Probe placement farther from the gastroesophageal junction may under-sample clinically relevant distal reflux events and contribute to symptom-physiology discordance. Observed disparities in reflux detection among female patients appear to be driven by differences in esophageal length rather than intrinsic physiological variation. These findings support prospective evaluation of anatomy-informed probe positioning to determine whether it can improve the consistency and diagnostic accuracy of ambulatory pH monitoring.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.