Impact of Photon-Counting CT Acquisition Modes on Coronary Stenosis Assessment and Radiation Dose: A Phantom and Clinical Evaluation.

Stein Thomas, Lang Friederike, Bamberg Fabian, Schlett Christopher L, Weiss Jakob, Ruile Philipp, Taron Jana

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Photon-counting detector computed tomography (PCD-CT) offers high spatial resolution and dose-efficient coronary CT angiography (cCTA), but protocol selection, especially optimal acquisition mode, requires balancing spacial resolution, temporal robustness and radiation exposure. This study evaluated various PCD-CT acquisition modes in terms of image quality, blooming, stenosis grading, diagnostic accuracy, and radiation dose using phantom and clinical cCTA.

METHODS

First-generation dual-source PCD-CT was used with four acquisition modes: high-pitch spiral with prospective ECG-gating (Flash), Sequence, Spiral, and retrospective ultra-high-resolution (UHR) spiral mode. Phantom experiments used six 3D-printed coronary artery models with stenoses ranging from 21% to 73%. In a retrospective clinical study, 41 cCTA examinations were analyzed. Qualitative and quantitative image assessments used invasive coronary angiography as reference standard for diagnostic accuracy. Patient characteristics and radiation doses were recorded.

RESULTS

UHR demonstrated the highest image quality and the lowest quantitative blooming in phantom and clinical data (clinical image quality score: UHR 4.12 ± 0.99 vs. Flash 3.00 ± 0.86; p = 0.002). Diagnostic accuracy varied across protocols for detecting ≥50% and ≥70% stenoses was high across all modes (e.g., Flash: 76.5% each; Sequence: 100% and 57.1%; UHR: 64.7% and 82.4%). Radiation dose was highest in UHR (26.44 ± 9.76 mSv) compared with Sequence (13.65 ± 8.86 mSv) and Flash (1.40 ± 0.32 mSv); all p ≤ 0.011.

CONCLUSION

While UHR acquisition provides highest image quality in PCD-CT cCTA, lower-dose acquisition modes achieve comparable diagnostic accuracy for coronary stenosis detection. Protocol selection should therefore be individualized, and prospective controlled studies are required.

Для цитирования: Stein Thomas, Lang Friederike, Bamberg Fabian, Schlett Christopher L, Weiss Jakob, Ruile Philipp, Taron Jana. Impact of Photon-Counting CT Acquisition Modes on Coronary Stenosis Assessment and Radiation Dose: A Phantom and Clinical Evaluation.. Zeitschrift fur medizinische Physik. 2026. DOI: 10.1016/j.zemedi.2026.10.001.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.