DOI 10.1016/j.pediatrneurol.2026.09.018

Медицинская библиотека · Pediatric neurology · 2026

Comparison of Ultrasound and Computed Tomography Optic Nerve Sheath Diameter for Intracranial Pressure Assessment in Pediatric Neurocritical Care.

García-Arellano Maricela, Hernández-Ortiz Deysi Guadalupe, Gaona-Espejo Jesús Alonso, Munguía-Chavez Sergio Erik, Marquez-Romero Juan Manuel

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Noninvasive intracranial pressure (ICP) assessment remains challenging in pediatric neurocritical care. Optic nerve sheath diameter (ONSD) measured by ultrasound (ONSD-US) and computed tomography (ONSD-CT) has been proposed as a surrogate marker of ICP, although comparative pediatric data remain limited.

METHODS

Prospective observational study conducted in a tertiary pediatric intensive care unit. Patients requiring invasive ICP monitoring were consecutively included. ONSD-US measurements were obtained simultaneously with invasive ICP measurements and, for patients undergoing CT, within 5 minutes before or after transfer for imaging. Pearson correlation and linear mixed-effects models accounting for repeated measurements were used. Bland-Altman analysis assessed intermodality agreement.

RESULTS

Eighteen patients contributed 139 measurement occasions. ONSD-US measurements were available in 116 occasions and ONSD-CT measurements in 37; 14 paired measurements from 11 patients included both modalities. One additional CT-derived measurement was excluded because of inadequate image quality, while the corresponding ONSD-US measurement was retained. ONSD-US showed a positive correlation with ICP (r = 0.32, P < 0.001), whereas ONSD-CT did not demonstrate a significant correlation (r = 0.23, P = 0.17). In mixed-effects analysis, each 1-mm increase in ONSD-US was associated with an estimated 3.24-mmHg increase in ICP (standard error 1.10; 95% confidence interval 1.09-5.39). Correlation between modalities was weak and not statistically significant (r = 0.31, P = 0.29).

CONCLUSIONS

ONSD-US demonstrated an association with invasively measured ICP after accounting for repeated measurements, whereas ONSD-CT did not show a clear association. Weak intermodality correlation suggests that ONSD-US and ONSD-CT should not be considered interchangeable methods for noninvasive ICP assessment in pediatric neurocritical care.

Для цитирования: García-Arellano Maricela, Hernández-Ortiz Deysi Guadalupe, Gaona-Espejo Jesús Alonso, Munguía-Chavez Sergio Erik, Marquez-Romero Juan Manuel. Comparison of Ultrasound and Computed Tomography Optic Nerve Sheath Diameter for Intracranial Pressure Assessment in Pediatric Neurocritical Care.. Pediatric neurology. 2026. DOI: 10.1016/j.pediatrneurol.2026.09.018.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.