Method Comparison of Three Automated D-Dimer Assays.
Аннотация
Background and
Objective
D-dimer results differ across analytical platforms, which may affect their clinical interpretation. This study aimed to compare the analytical agreement and threshold-based classification of three routinely used D-dimer analyzers.
Methods
Of 150 consecutive residual citrated plasma specimens, 81 patient specimens with valid results on the Sysmex CS-2500, Tokra Novae MTI, and VIDAS 3 analyzers were included. Agreement was assessed using the intraclass correlation coefficient, Passing-Bablok regression, Bland-Altman analysis, and classification concordance at the fixed (500 ng/mL FEU) and age-adjusted thresholds. VIDAS 3 served solely as the comparator method.
Results
Overall agreement was moderate to good; however, the platforms were not fully interchangeable. Sysmex CS-2500 and VIDAS 3 showed the closest agreement, with no significant constant or proportional bias, whereas Tokra Novae MTI exhibited significant proportional bias. Mixed classification across the three analyzers occurred in 21/81 patients (25.9%) at the fixed threshold and in 18/81 patients (22.2%) at the age-adjusted thresholds.
Conclusions
Clinically relevant inter-platform differences persisted despite acceptable average agreement. Therefore, D-dimer results should not be regarded as interchangeable across analyzers, particularly near clinical decision thresholds, and platform-specific verification remains essential.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.