MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.70948

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026

Catheter-Directed Interventional Devices for Pulmonary Embolism: A Maude Database Analysis of Adverse Events.

Nandyal Shreyas, Gajjar Rohan, Varma Revati, Vinjamuri Saketh, Athmakuri Krishna Saketh, Vij Aviral

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Catheter-directed therapies for acute pulmonary embolism (PE) are being adopted rapidly, yet real-world safety data remain limited.

AIMS: To characterize adverse events, device failure patterns, and mortality signals for three FDA-cleared catheter-directed PE platforms (EKOSONIC, INDIGO, and FlowTriever) using the FDA Manufacturer and User Facility Device Experience (MAUDE) database.

METHODS

Medical device reports (MDRs) were downloaded from each device's PE clearance (2008, 2015, and 2018, respectively) through March 2026 and filtered by text to retain PE-specific reports. Reports were classified by manufacturer, event type (death, injury, malfunction), device problem, and patient problem, and trends were assessed by calendar year.

RESULTS

Of 2528 reports, 389 were PE-specific (INDIGO 228 [58.6%], EKOSONIC 149 [38.3%], FlowTriever 12 [3.1%]), spanning 2009 to January 2026. There were 76 deaths (19.5%), 133 injuries (34.2%), and 180 malfunctions (46.3%). Annual death reports rose from 2023 (15 in 2023, 13 in 2024, 8 in 2025, and 3 in January 2026). Among 47 INDIGO deaths, 28 (59.6%) involved Lightning/Flash large-bore catheters, and 36 (76.6%) carried a perforation-, hemoptysis-, tamponade-, or extravasation-related patient problem code. Separator wire break/fracture was the leading INDIGO device failure. EKOSONIC reports showed thrombolytic-related bleeding, including intracranial or cerebral hemorrhage in 12 reports (8.1%).

CONCLUSIONS

MAUDE reports reveal distinct device-specific safety signals: pulmonary artery perforation with large-bore INDIGO aspiration catheters, and intracranial hemorrhage with EKOSONIC. Death reports increased from 2023 as mechanical thrombectomy use expanded. Absent procedural denominators, rates and causality cannot be inferred; findings support operator awareness, careful patient selection, and continued post-market surveillance.

Для цитирования: Nandyal Shreyas, Gajjar Rohan, Varma Revati, Vinjamuri Saketh, Athmakuri Krishna Saketh, Vij Aviral. Catheter-Directed Interventional Devices for Pulmonary Embolism: A Maude Database Analysis of Adverse Events.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2026. DOI: 10.1002/ccd.70948.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.