Medical engineering & physics · 2026
Effects of LBBAP Pacing-Site Position under a Modelled Proximal LBBB: A 24-Model Electrophysiological Simulation Study.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Tang Biao, Pei Zhiqiang, Zhang Tianyi, Zhai Qichao, Cheng Yunzhang
Аннотация
Left bundle branch area pacing (LBBAP) may recruit the distal left-sided conduction 
system, but the electrical effect of pacing-site position along the left bundle branch (LBB) 
remains incompletely defined. We evaluated this effect in 24 four-chamber virtual heart 
anatomies using a Reaction-Eikonal finite-element model with rule-based myocardial 
fibres and algorithmically generated His-Purkinje networks. A modelled proximal His
to-LBB disconnection preserved the distal LBB-Purkinje network while interrupting its 
proximal connection. Six conditions were simulated in each anatomy: a sinoatrial-node 
(SAN)-initiated left bundle branch block (LBBB) control, right ventricular apical pacing, 
right ventricular outflow-tract pacing and a modelled non-selective LBB capture 
configuration at proximal, nominal and distal positions. The primary outcome was 
BiV_TAT95, the 95th percentile of activation times across left-ventricular, right
ventricular and interventricular-septal nodes. The control-versus-paced analysis used a 
common simulation-clock origin and was interpreted as contextual benchmarking, 
whereas the positional analysis used matched direct ventricular stimulation. In the 
matched positional analysis, median BiV_TAT95 was 104.5, 93.5 and 88.5 ms for 
proximal, nominal and distal LBBAP, respectively (Friedman chi-square = 48.000, P < 
0.001, Kendall's W = 1.000); every anatomy followed distal < nominal < proximal. In 
three selected anatomies, a 20% decrease or increase in the prescribed myocardial 
conduction speeds altered absolute activation times but preserved this ordering in all 
nine anatomy-by-speed combinations. Under the tested homogeneous substrate and 
modelled non-selective capture configuration, the distal tested position produced the 
lowest simulated BiV_TAT95; this does not imply superiority for every activation metric. 
This conditional prediction requires validation against patient-specific electrical data.