MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.eplepsyres.2026.107933

Медицинская библиотека · Epilepsy research · 2026

Impact of entry trajectory angle-to-skull on electrode placement accuracy in robot-assisted stereoelectroencephalography.

Park Chan, Kim Jong Hyun, Kim Ji Hyun, Shin Ji Hye, Roh Haewon

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Despite the widespread use of robotic systems to improve stereotactic accuracy in stereoelectroencephalography (SEEG), targeting errors still occur. Several studies have suggested that the trajectory-to-skull angle is closely associated with targeting error among various influencing factors; however, quantitative analyses remain limited. Therefore, this study aimed to quantitatively evaluate the relationship between trajectory-to-skull angle and implantation accuracy and to establish practical cut-off values for surgical planning.

METHODS

A total of 10 patients (103 electrodes) were included in this study and various clinical variables were analyzed to identify whether trajectory-to-skull angle has a significant relation with accuracy using multivariate linear regression. Then, receiver operating characteristic (ROC) curve analysis with Youden index optimization was applied to determine angle thresholds for achieving ≤ 3 mm accuracy.

RESULTS

Multivariate analysis showed that both axial and coronal trajectory-to-skull angles were significantly associated with target error, with more tangential incidence angles corresponding to larger deviations (p < 0.01). Axial angle-to-skull was particularly associated with anteroposterior (Y-axis) error, whereas coronal angle-to-skull primarily predicted superoinferior (Z-axis) error. ROC curve analysis identified optimal angle thresholds of 11.8° for axial trajectories and 6.3° for coronal trajectories to predict targeting accuracy within 3 mm. When both axial and coronal angles were maintained within these cut-offs, the probability of achieving a target error ≤ 3 mm was 88.2%.

CONCLUSION

Trajectory-to-skull angle is a key geometric determinant of electrode placement accuracy in robot-assisted SEEG. Quantitative angle thresholds identified in this study provide practical guidance for trajectory planning and may help minimize mechanical deviation, improve implantation safety, and enhance the reliability of seizure-network localization.

Для цитирования: Park Chan, Kim Jong Hyun, Kim Ji Hyun, Shin Ji Hye, Roh Haewon. Impact of entry trajectory angle-to-skull on electrode placement accuracy in robot-assisted stereoelectroencephalography.. Epilepsy research. 2026. DOI: 10.1016/j.eplepsyres.2026.107933.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.