Kang Le, Cui Jian-Qiang, Dong Yu-Jin, Guo Bao-Lei, Qin Xue-Meng, Yang Mao-Wei
Аннотация
OBJECTIVE: To develop an individualised distal fibular locking hook plate for Danis-Weber type A fractures and assess its preclinical biomechanics.
METHODS: Anatomical measurements of the anterior talofibular ligament (ATFL) and calcaneofibular ligament (CFL) were pooled to define ligament attachment sites and optimise hook-tooth placement. A cobalt-chromium alloy locking hook plate was designed using computer-aided design software, fabricated by 3D printing, and evaluated by finite element analysis under static, inversion, eversion, and external-rotation loading conditions.
RESULTS: A 9.0-mm gap between the hook teeth was selected from the pooled anatomical measurements. Across all tested loading conditions, fracture-site displacement remained below 1.0 mm. Under 700-N external-rotation loading, the maximum von Mises stress was 23.0 MPa in the distal fibula and 53.9 MPa in the plate.
CONCLUSION: The plate demonstrated anatomical compatibility and favourable finite element performance. Cadaveric testing and prospective clinical studies are required before clinical effectiveness or routine use can be established.
METHODS: Anatomical measurements of the anterior talofibular ligament (ATFL) and calcaneofibular ligament (CFL) were pooled to define ligament attachment sites and optimise hook-tooth placement. A cobalt-chromium alloy locking hook plate was designed using computer-aided design software, fabricated by 3D printing, and evaluated by finite element analysis under static, inversion, eversion, and external-rotation loading conditions.
RESULTS: A 9.0-mm gap between the hook teeth was selected from the pooled anatomical measurements. Across all tested loading conditions, fracture-site displacement remained below 1.0 mm. Under 700-N external-rotation loading, the maximum von Mises stress was 23.0 MPa in the distal fibula and 53.9 MPa in the plate.