A patient contact shield visible in the localizer-the effects to tube current modulation.

Larjava Heli R S, Eneh Chibuzor Toivo Mark, Niskanen Eini, Järvinen Lauri, Sierpowska Joanna, Teerimäki Päivi, Järvinen Jukka, Saunavaara Virva, Huuskonen Minna, Niiniviita Hannele

СтатьяПеревод пока не подготовлен

Аннотация

An anthropomorphic phantom was scanned with Siemens Somatom computed tomography scanners using tube current modulation (TCM). Thorax of the phantom was first scanned without the use of a contact shield, followed by a scan with a 0.5 mm lead equivalent out-of-plane shield placed in the localizer but 1 cm caudally outside the scanned volume. Scans were repeated using the scanner's organ radiation dose reduction feature. Pitch values 0.35, 0.6, and 1.05 were used. With a pitch of 1.05, the effects of TCM varied between scanners from 4.35% to 356% when scanning toward the caudal end of the scanned volume, when shield was used outside the scanned volume. Also, with some scanners, the use of XCare increased the cumulative tube current.

Для цитирования: Larjava Heli R S, Eneh Chibuzor Toivo Mark, Niskanen Eini, Järvinen Lauri, Sierpowska Joanna, Teerimäki Päivi, Järvinen Jukka, Saunavaara Virva, Huuskonen Minna, Niiniviita Hannele. A patient contact shield visible in the localizer-the effects to tube current modulation.. Radiation protection dosimetry. 2026. DOI: 10.1093/rpd/ncag125.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.