Independent Three-dimensional Dose Calculation from Log Files for Breast VMAT: A Measurement-free Quality Assurance Approach.

Sevilla Rosa Angelica Petit, Blanco Oscar Abel Apaza, Soares Nayron Medeiros, Musso Maximiliano, Lacey Ezequiel Agustín Cabrera, Bohorquez Luis C, Wolfsberger Lucy, Almada Maria José, Venencia Carlos Daniel

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The purpose of this study was to evaluate and validate a measurement-free patient-specific quality assurance (PSQA) approach for volumetric modulated arc therapy (VMAT) breast cancer treatments based on independent three-dimensional (3D) dose calculation from Digital Imaging and Communications in Medicine (DICOM) plans and linear accelerator log files.

MATERIALS AND

METHODS

Seventy clinical breast VMAT plans (36 left-sided and 34 right-sided) were retrospectively analyzed. Independent 3D dose calculations were performed using RadCalc v7.3 based on the DICOM treatment plan (independent dose calculation [IDC] RTPlan) and on delivery log files acquired during the first treatment fraction (IDC LOGs). Results were compared with portal dosimetry (two-dimensional [2D]) and fluence-based analysis. Gamma passing rates (GPRs) were evaluated using 3%/2 mm and 2%/2 mm criteria under action (90%) and tolerance (95%) limits. Receiver operating characteristic (ROC) analysis was performed to evaluate the discriminatory capability. Sensitivity to delivery errors was investigated by introducing controlled multileaf collimator (MLC) shifts (0.5-2.0 mm) and collimator rotation errors (1°-3°).

RESULTS

All PSQA methods met the TG-218 action limit (GPR ≥ 90%) at 3%/2 mm. Under the stricter 2%/2 mm criterion, statistically significant differences among quality assurance methods were observed (P < 0.001). ROC analysis demonstrated superior discriminatory performance for 3D dose-based methods, with area under the curve values of 0.9 for both IDC RTPlan and IDC LOGs, compared with 0.7 for portal dosimetry and 0.5 for fluence-based analysis. Error sensitivity analysis showed that stringent gamma criteria (1%/2 mm and 1%/1 mm) improved detection of subtle MLC and collimator deviations, with log file-based 3D reconstruction outperforming 2D approaches.

CONCLUSIONS

Independent 3D dose calculation is a reliable PSQA approach for breast VMAT, showing higher sensitivity than conventional 2D methods.

Для цитирования: Sevilla Rosa Angelica Petit, Blanco Oscar Abel Apaza, Soares Nayron Medeiros, Musso Maximiliano, Lacey Ezequiel Agustín Cabrera, Bohorquez Luis C, Wolfsberger Lucy, Almada Maria José, Venencia Carlos Daniel. Independent Three-dimensional Dose Calculation from Log Files for Breast VMAT: A Measurement-free Quality Assurance Approach.. Journal of medical physics. 2026. DOI: 10.4103/jmp.jmp_38_26.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.