Prospective controlled ex vivo observational study on changes in elasticity of ureteral stents following clinical use.

Hoffmann Marina, Ebner Benedikt, Pyrgidis Nikolaos, Blajan Iulia, Georgieva Troya, Kidess Marc, Happe Moritz, Volz Yannic, Eismann Lennert, Apfelbeck Maria, Stief Christian G, Chaloupka Michael

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Ureteral stents significantly impair health-related quality of life. Advances in stent material, particularly in polyurethane compositions, may mitigate adverse effects. This study evaluated ex vivo changes in the elasticity of ureteral stents made from a novel thermoplastic polyurethane composite (HybridPRO Technology by Urotech) compared to a control stent.

MATERIAL AND

METHODS

In this prospective controlled ex vivo observational study, 97 patients undergoing ureteroscopy received one of two types of tethered 7 Fr, 28 cm ureteral stents. Eleven patients underwent bilateral procedures, yielding a total of 108 stents (54 per group). Group 1 received a polyurethane stent with HybridPRO Technology and group 2 received a polyurethane stent without HybridPRO Technology. After removal, stents were measured at the patient's bedside using a hanging apparatus with a predetermined 400 g load. In vitro (room temperature) and ex vivo elongations were recorded, and the net difference was calculated.

RESULTS

In vitro measurements showed a median elongation of 4 mm (IQR: 4-4) in group 1 vs 2 mm (IQR: 2-2) in group 2. Ex vivo, group 1 elongated significantly more (median, 12 mm; IQR: 11-13) compared to group 2 (median: 5 mm; IQR: 4-6; p <0.001). The net difference was larger in group 1 (median: 10 mm; IQR, 9-11) compared to group 2 (median: 3 mm; IQR: 2-4; p <0.001).

CONCLUSIONS

Ureteral stents incorporating HybridPRO Technology demonstrated significantly greater elongation under standardized load after use compared to a control stent. These findings indicate material changes ex vivo; however, the clinical implications regarding comfort or quality of life remain to be determined in future studies.

Для цитирования: Hoffmann Marina, Ebner Benedikt, Pyrgidis Nikolaos, Blajan Iulia, Georgieva Troya, Kidess Marc, Happe Moritz, Volz Yannic, Eismann Lennert, Apfelbeck Maria, Stief Christian G, Chaloupka Michael. Prospective controlled ex vivo observational study on changes in elasticity of ureteral stents following clinical use.. Central European journal of urology. 2026. DOI: 10.5173/ceju.2025.0216.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.