MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcjo.2025.02.019

Медицинская библиотека · Canadian journal of ophthalmology. Journal canadien d'ophtalmologie · 2025

Magnetic resonance imaging in preseptal ferromagnetic foreign bodies.

Ma Tu-An, Rana Khizar, Tong Jessica Y, Ullrich Katja, Patel Sandy, Selva Dinesh

СтатьяКлинический случайОбзорПеревод в очереди

Аннотация

OBJECTIVE

To review the literature on safety of magnetic resonance imaging (MRI) in patients with preseptal ferromagnetic foreign bodies.

METHODS

We describe 3 cases of MRI in patients with preseptal ferromagnetic foreign bodies (FFBs) from our institution.

RESULTS

The FFBs were all preseptal, adjacent to the medial canthus (n = 2) and lateral canthus (n = 1). None of the patients had any ocular complications post-MRI. The literature review identified an additional 7 cases with intraocular and preseptal FFBs that underwent MRI. The FFBs ranged in size from 1.0 mm to 3.5 mm. The FFBs were intraocular (n = 6), or preseptal (n = 1). The MRI field strength ranged from 0.35 T to 1.5 T. Five (83.3%) of the patients with intraocular FFBs had ocular complications, which included hyphema (n = 2), cataract (n = 3), vitreous haemorrhage (n = 1), and corneal scar (n = 1). The patient with preseptal FFB did not have post-MRI complications.

CONCLUSIONS

There is some evidence to suggest that patients with preseptal FFBs may be less likely to experience complications post-MRI compared to intraocular FFBs. Various factors affect the safety of MRIs, including FFB location, size, proximity to visually-significant structures, and MRI field strength.

Для цитирования: Ma Tu-An, Rana Khizar, Tong Jessica Y, Ullrich Katja, Patel Sandy, Selva Dinesh. Magnetic resonance imaging in preseptal ferromagnetic foreign bodies.. Canadian journal of ophthalmology. Journal canadien d'ophtalmologie. 2025. DOI: 10.1016/j.jcjo.2025.02.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.