A Rare Cause of Stapedotomy Failure: When the Depth Marker Becomes a Locking Device: A Report on 2 Cases.

Daval Mary, Guevara Nicolas, Alotaibi Naif, Williams Marc, Ayache Denis

Клинический случайСтатьяПеревод в очереди

Аннотация

We report 2 cases of stapes surgery failure due to the depth marker groove of the piston of a Big Easy titanium prosthesis getting caught at the edge of the stapedotomy hole. In the first case, the stapedotomy failure, detected peroperatively, was corrected by modifying the angulation of the shaft of the prosthesis. In the second case, the failure occurred after a delay during which scar tissue developed around the depth marker groove, hampering the movement of the prosthesis. The usual postoperative imaging did not detect this complication. The diagnosis was made during revision surgery and the condition was treated by enlarging the hole in the footplate. Alternatively, the use of a piston without a depth marker groove would have resolved the problem. Although a rare event, the possibility of stapes surgery failure due to the depth marker groove should not be overlooked.

Для цитирования: Daval Mary, Guevara Nicolas, Alotaibi Naif, Williams Marc, Ayache Denis. A Rare Cause of Stapedotomy Failure: When the Depth Marker Becomes a Locking Device: A Report on 2 Cases.. Ear, nose, & throat journal. 2019. DOI: 10.1177/0145561319839869.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.