Huang Po-Kai, Chen Chin-Kuo, Yu Yi-Hsuan, Ho Guan-Min, Hsieh Li-Chun
Аннотация
OBJECTIVES: Medialization thyroplasty (MT) using various implants has been employed as a corrective procedure for unilateral vocal fold paralysis (UVFP). A newly developed APrevent® vocal implant system (VOIS) offers an innovative solution with a finely adjustable design. This study aimed to investigate the long-term functional voice outcomes and benefits of postoperative adjustments in patients receiving MT using the VOIS-implant.
METHODS: This is a prospective case series study at single tertiary medical center. Fourteen adult patients diagnosed with UVFP received MT with the VOIS implant and were followed up for more than 1 year. Implant adjustment procedure by injecting 0.9% physiological saline solution was performed both during and after the surgery to optimize glottal closure and voice quality. Objective voice outcomes and acoustic parameters were assessed preoperatively and postoperatively at various timepoints.
RESULTS: Thirteen patients (93%) received intraoperative balloon adjustment, ranging from 0.05to 0.12 ml. Four patients underwent adjustments postoperatively and exhibited a positive trend towards immediately improving acoustic voice quality. Our long-term results demonstrated a notable improvement after the surgery in voice quality, with significant decreases in VHI-30 and improvements in perceptual parameters of GRBAS scale, acoustic measures such as jitter and signal-to-noise ratio (p < 0.001) and cepstral peak prominence smoothed in sustained vowel and short sentences. The voice outcomes remained stable more than 1 year follow-up.
CONCLUSIONS: Overall, MT with VOIS implantation provides a favorable long-term outcomes and stability in voice quality for patients with UVFP and also an effective tool for postoperative adjustment without major revision surgeries.
METHODS: This is a prospective case series study at single tertiary medical center. Fourteen adult patients diagnosed with UVFP received MT with the VOIS implant and were followed up for more than 1 year. Implant adjustment procedure by injecting 0.9% physiological saline solution was performed both during and after the surgery to optimize glottal closure and voice quality. Objective voice outcomes and acoustic parameters were assessed preoperatively and postoperatively at various timepoints.
RESULTS: Thirteen patients (93%) received intraoperative balloon adjustment, ranging from 0.05to 0.12 ml. Four patients underwent adjustments postoperatively and exhibited a positive trend towards immediately improving acoustic voice quality. Our long-term results demonstrated a notable improvement after the surgery in voice quality, with significant decreases in VHI-30 and improvements in perceptual parameters of GRBAS scale, acoustic measures such as jitter and signal-to-noise ratio (p < 0.001) and cepstral peak prominence smoothed in sustained vowel and short sentences. The voice outcomes remained stable more than 1 year follow-up.
CONCLUSIONS: Overall, MT with VOIS implantation provides a favorable long-term outcomes and stability in voice quality for patients with UVFP and also an effective tool for postoperative adjustment without major revision surgeries.