[Accuracy of home-use self-rebound-tonometer for intraocular pressure measurement in glaucoma and non-glaucoma patients].
Аннотация
Objective
To compare the measurement results of a home-use self-tonometer and the Goldmann applanation tonometer (GAT) in glaucoma and non-glaucoma subjects, evaluate the accuracy of the home-use self-tonometer, and investigate potential factors influencing the measurement outcomes.
Methods
It was a cross-sectional study. Non-glaucoma and glaucoma subjects were consecutively recruited from the Glaucoma Clinic of the Zhongshan Ophthalmic Center, Sun Yat-sen University, between August 2025 and November 2025. After receiving standardized training and certification for the home-use self-tonometer, subjects underwent three sequential intraocular pressure (IOP) measurements: self-measurement by the subject, measurement by a physician (using the home tonometer), and GAT measurement. Differences, correlations, and agreement among the three IOP measurements were analyzed using paired one-way analysis of variance (ANOVA), Pearson correlation coefficients, and Bland-Altman analysis with 95% limits of agreement. Paired t-tests were used to analyze whether there were differences in measurement errors between the left and right eyes during subject self-measurement. Univariate and multivariate regression analysis were performed to evaluate factors associated with measurement errors of the home-use self-rebound tonometer Both eyes were included when analyzing the difference in measurement errors between left and right eyes; for all other statistical analyses, only data from one eye per subject were included.
Results
A total of 69 subjects (132 eyes) were recruited, with a mean age of (38.16±11.57) years, including 15 non-glaucoma subjects (30 eyes) and 54 glaucoma patients (102 eyes). Sixty-six subjects (95.65%) passed the certification. The IOP values obtained from subject self-measurement, physician measurement, and GAT were (15.74±4.40) mmHg, (15.70±5.22) mmHg (1 mmHg=0.133 kPa), and (15.88±5.10) mmHg, respectively, with no statistically significant differences among the three methods (F=0.12, P=0.874). Strong correlations were observed between GAT measurements and physician measurements, between GAT measurements and subject self-measurements, and between physician measurements and subject self-measurements (r=0.86, 0.75, and 0.78, respectively; all P<0.001). Using GAT values as the reference, the mean biases between GAT and physician measurements and between GAT and subject self-measurements were (0.19±2.70) mmHg and (0.14±3.41) mmHg, respectively, while the mean bias between the physician and the subject using the home-use self-tonometer was (-0.05±3.27) mmHg. Compared with GAT measurements, the measurement errors of subject self-measurement in the right and left eyes were (-0.05±3.21) mmHg and (-0.03±2.72) mmHg, respectively, with no significant difference between eyes (P=0.872). In univariate linear regression analysis, the measurement error of subject self-measurement was positively associated with GAT-measured IOP (β=0.35, 95%CI: 0.21-0.49, P<0.001). The measurement differences between physician and subject measurements using the home-use self-rebound tonometer were positively associated with best-corrected visual acuity (BCVA) and GAT-measured IOP (β=5.26 and 0.24, respectively; 95%CI: 2.04-8.48 and 0.09-0.38, respectively; both P=0.002). Glaucoma was associated with increased measurement differences between physician and subject measurements using the home-use self-rebound tonometer (β=2.10, 95%CI: 0.27-3.92, P=0.028). In multivariate linear regression analysis, GAT-measured IOP was an independent factor associated with physician measurement error (β=-0.29, 95%CI: 0.06-0.55, P=0.022).
Conclusion
The home-use self-rebound-tonometer showed good agreement with the Goldmann applanation tonometer in both glaucoma and non-glaucoma subjects, indicating that the measurement results obtained using the home-use self-rebound tonometer are reliable.