Laser-assisted fixation of a nitinol stapes prosthesis.
Аннотация
BACKGROUND AND
OBJECTIVE
Otosclerosis is an inner ear bone disease characterized by fixation of the stapes and consequently progressive hearing loss. One treatment option is the surgical replacement of the stapes by a prosthesis. When so called "smart materials" like nitinol are used, prosthesis fixation can be performed using a laser without manual crimping on the incus. However, specific laser-prosthesis interactions have not been described yet. The aim of the present study was to elucidate the thermo-mechanical properties of the NiTiBOND® prosthesis as a basis for handling instructions for laser-assisted prosthesis fixation.
STUDY DESIGN AND MATERIALS AND
METHODS
Closure of the NiTiBOND® prosthesis was induced ex vivo by either a diode laser emitting at λ = 940 nm or a CO2 laser (λ = 10,600 nm). Total energy for closure was determined. Suitable laser parameters (pulse duration, power per pulse, distance between tip of the laser fiber and prosthesis) were assessed. Specific laser-prosthesis interactions were recorded.
RESULTS
Especially the diode laser was found to be an appropriate energy source. A total energy deposit of 60 mJ by pulses in near contact application was found to be sufficient for prosthesis closure ex vivo. Energy should be transmitted through a laser fiber equipollent to the prosthesis band diameter. Specific deformation characteristics due to the zonal prosthesis composition have to be taken into account.
CONCLUSION
NiTiBOND® stapes prosthesis can be closed by very little energy when appropriate energy sources like diode lasers are used, suggesting a relatively safe application in vivo. Lasers Surg. Med. 50:153-157, 2018. © 2017 Wiley Periodicals, Inc.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.