MedElevator Войти в исследования
← Поиск публикаций

DOI 10.7874/jao.2017.21.2.103

Медицинская библиотека · Journal of audiology & otology · 2017

Comparison of Free-Beam- and Fiber-Type CO2 Laser Delivery Systems in Stapes Surgery.

Chang Mun Young, Choi Hyun Seok, Lee Sang-Youp, Koo Ja-Won

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND

OBJECTIVES

A free-beam-type CO2 laser, which use a micromanipulator mounted on a microscope as the delivery system, has the merit of not being affected by hand tremor at the time of shooting. However, this delivery system has several disadvantages, including a restricted operation range and a risk of incorrect focusing. A fiber-type CO2 laser uses a hand-held delivery system and has the opposite merits and demerits. We compared the results of stapes surgery with free-beam and fiber type delivery systems.

SUBJECTS AND

METHODS

The study enrolled 36 patients who underwent stapedotomy with free-beam- (n=26) or fiber- (n=10) type CO2 lasers. The air-bone (AB) gap closure, bone conduction (BC) change, and operating time were evaluated. The AB gap closure was calculated by subtracting the preoperative BC thresholds from the postoperative air conduction thresholds. The BC change was calculated by subtracting the postoperative BC thresholds from the preoperative BC thresholds.

RESULTS

The mean operating time was significantly (p=0.035) shorter in the fiber-type group (72.5±8.2 min) than in the free-beam-type group (80.5±11.4 min). The mean AB gap closure did not differ significantly (p=0.297) between the free-beamand fiber-type groups (5.8±10.1 and 1.4±6.8 dB, respectively). The mean BC change did not differ significantly (p=0.873) between the free-beam- and fiber-type groups (2.4±6.9 and 2.8±5.3 dB, respectively). The hearing outcomes did not differ significantly between the two groups.

CONCLUSIONS

Operating times were significantly shorter using the fiber-type CO2 laser, while hearing outcomes did not differ significantly between the two groups.

Для цитирования: Chang Mun Young, Choi Hyun Seok, Lee Sang-Youp, Koo Ja-Won. Comparison of Free-Beam- and Fiber-Type CO2 Laser Delivery Systems in Stapes Surgery.. Journal of audiology & otology. 2017. DOI: 10.7874/jao.2017.21.2.103.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.