Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.
Аннотация
BACKGROUND
Analytical quality of blood glucose monitoring systems (BGMS) is an important aspect for many diabetes patients. Sufficiently high analytical quality is required for adequate diabetes therapy.
METHODS
In this study, system accuracy and measurement precision of a BGMS were assessed based on ISO 15197:2013. For system accuracy, this standard requires a specific glucose distribution and at least 95% of results obtained with the BGMS in capillary blood to fall within ±15 mg/dl or ±15% (at glucose concentrations <100 mg/dl or ≥100 mg/dl, respectively) of corresponding comparison method results, and at least 99% of results to be found within clinically acceptable consensus error grid (CEG) zones A and B. Based on ISO 15197:2013, intermediate measurement precision, using control solution, and measurement repeatability, using venous blood samples, were analyzed by calculation of standard deviations (SDs) and coefficients of variation (CV) at glucose concentrations <100 mg/dl or ≥100 mg/dl, respectively, although ISO 15197:2013 does not specify acceptance criteria.
RESULTS
The BGMS fulfilled system accuracy requirements with ≥99% of results within ±15 mg/dl or ±15% of the comparison method results, and 100% of results in CEG zones A and B. Intermediate measurement precision analysis showed SD ≤2.2 mg/dl and CV ≤2.3%. Analysis of measurement repeatability showed SD ≤2.1 mg/dl and CV ≤2.4%.
CONCLUSION
System accuracy requirements of ISO 15197:2013 were fulfilled by the BGMS. As ISO 15197:2013 does not specify precision requirements, precision analysis results were compared with those reported for other BGMS in the literature and found to be similar.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.