Cardiovascular revascularization medicine : including molecular interventions · 2025

Effect of optimal lesion preparation prior to implantation of a magnesium-based bioresorbable scaffold: 12-month results of the OPTIMIS study.

СтатьяРандомизированное исследованиеСравнительное исследованиеМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Hansen Kirstine Nørregaard, Maehara Akiko, Trøan Jens, Noori Manijeh, Hougaard Mikkel, Ellert Julia, Veien Karsten Tange, Junker Anders, Hansen Henrik Steen, Lassen Jens Flensted, Jensen Lisette Okkels

Аннотация

BACKGROUND: Lumen reduction after bioresorbable scaffold implantation has been reported. This study aimed to assess the influence of pre-dilatation with a scoring balloon versus a standard non-compliant balloon prior to implanting a magnesium-based Magmaris bioresorbable scaffold (MgBRS) on lumen measurements using optical coherence tomography (OCT) and on clinical outcomes after 12 months.

METHOD: In the OPTIMIS-study (Optimal lesion preparation before implantation of a MgBRS in patients with coronary artery stenosis), patients were randomly assigned to pre-dilatation with a scoring balloon or a standard non-compliant balloon before MgBRS implantation. OCT was performed before and after scaffold implantation, and at 6- and 12-month follow-up. Clinical 12-month follow-up was performed for all 82 enrolled patients.

RESULTS: Serial pre- and post-procedural, 6- and 12-month OCT were available in 33 lesions (scoring: n = 16 vs. standard: n = 17). Minimal lumen area (MLA) (scoring 4.6 ± 0.5 mm2 vs. standard 3.6 ± 0.5 mm2, p = 0.16) did not differ significantly at 12 months. In both the scoring and standard non-compliant balloon group, MLA changed significantly over time (post-procedure - 6-month - 12-month) (scoring balloon: 7.2 ± 1.8 mm2-5.1 ± 1.7 mm2-5.0 ± 1.9 mm2, p < 0.01, and standard non-compliant balloon: 6.2 ± 1.1 mm2-5.2 ± 1.5 mm2-5.3 ± 2.0 mm2, p < 0.01). No target lesion revascularizations (TLR) occurred in the scoring balloon group, compared to 7 TLR in the standard non-compliant balloon group within 12 months.

CONCLUSION: In lesions treated with MgBRS, MLA did not differ significantly after 12 months among lesions pre-dilated with a scoring balloon compared to a standard non-compliant balloon. However, only lesions pre-dilated with a scoring balloon were free from TLR.