MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3389/fneur.2026.1781522

Медицинская библиотека · Frontiers in neurology · 2026

Investigation of distributive, morphological and audiological factors influencing the long-term aggravation of tinnitus following cochlear implant provision.

Everad Friederike, Aschendorff Antje, Rauch Ann-Kathrin, Fries Leonie, Arndt Susan, Ketterer Manuel Christoph

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

This prospective study aims to identify factors that may influence the long-term aggravation of tinnitus following cochlear implant (CI) surgery. The variables examined include age, gender, pre- and post-operative residual hearing, duration of deafness, etiology of hearing impairment, cochlear anatomy, as well as the insertion depth and angle of the CI electrode array.

METHODS

A total of 65 patients were assessed pre-operatively and at 2 days, 4 weeks, 12 months, and 24 months after CI surgery. Age, gender, duration of deafness, and etiology were recorded through anamnesis. Residual hearing before and after implantation was evaluated using the air conduction (AC) thresholds in four frequencies (500, 1,000, 2,000, and 4,000 Hz) in pure-tone audiometry (PTA-4). Cochlear dimensions (distances and height), insertion depth, and insertion angle were measured using post-operative digital volume tomography. Tinnitus burden and intensity were assessed with the Visual Analog Scale (VAS) and the Tinnitus Questionnaire.

RESULTS

Overall, tinnitus burden significantly decreased during the course of CI treatment. However, 38% of patients reported increased tinnitus intensity on the second post-operative day, with this proportion remaining relatively stable (36%) after 2 years. Meanwhile, 33% of patients experienced an increase of tinnitus burden shortly after implantation, decreasing to 21% 1 year post-operatively. None of the examined factors-including age, gender, residual hearing, duration of deafness, etiology, cochlear anatomy, insertion depth, or insertion angle-were associated with an increased risk of pre-operative tinnitus burden or post-operative tinnitus exacerbation.

CONCLUSION

Distributive and morphological factors did not significantly influence tinnitus exacerbation before or after CI. Nevertheless, tinnitus symptoms improved significantly over time. Future studies should investigate additional potential contributing factors, such as psychological comorbidities, in the development or persistence of tinnitus following CI.

Для цитирования: Everad Friederike, Aschendorff Antje, Rauch Ann-Kathrin, Fries Leonie, Arndt Susan, Ketterer Manuel Christoph. Investigation of distributive, morphological and audiological factors influencing the long-term aggravation of tinnitus following cochlear implant provision.. Frontiers in neurology. 2026. DOI: 10.3389/fneur.2026.1781522.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.