Early outcomes of Myval™ Octacor transcatheter pulmonary valve implantation in dysfunctional right ventricular outflow tract conduits and pulmonary valve bioprostheses: a single-centre UK experience.
Аннотация
BACKGROUND
Transcatheter pulmonary valve implantation has emerged as a minimally invasive and preferred therapeutic option for patients with dysfunction of previously repaired right ventricular outflow tracts. The Myval™ Octacor valve is a new device designed for this purpose, though limited reports exist regarding its use in the pulmonary position.
AIMS: To report the immediate and short-term outcomes of percutaneous pulmonary valve implantation using the Myval™ Octacor valve in patients with severe right ventricular-pulmonary artery conduit or pulmonary valve bioprosthesis dysfunction.
METHODS
This was a single-centre retrospective review of data obtained from case files.
RESULTS
The Myval™ Octacor valve was used in ten patients with a mean age of 34.5 ± 7.4 years. The median procedure duration and fluoroscopy time were 146 minutes and 30.5 minutes, respectively. The median Z-score for valves used was -0.5. The median right ventricular systolic pressure decreased from 68.5 mmHg pre-procedure to 33 mmHg post-procedure. The median peak instantaneous gradient across the right ventricular outflow tract or conduit decreased from 30 mmHg to 6.5 mmHg. There were no reported incidences of frame fracture, conduit rupture, device embolisation, or endocarditis.
CONCLUSION
This is the first UK experience of using the new-generation Myval™ Octacor valve in percutaneous pulmonary valve implantation. The results demonstrate the valve's safety and clinical efficacy, with favourable outcomes in terms of procedural success, haemodynamic improvement, and echocardiographic findings.