DOI 10.1016/j.jacc.2006.03.037

Медицинская библиотека · Journal of the American College of Cardiology · 2006

A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial.

Wilkoff Bruce L, Ousdigian Kevin T, Sterns Laurence D, Wang Zengri J, Wilson Ryan D, Morgan John M, EMPIRIC Trial Investigators

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The purpose of this randomized study was to determine whether a strategically chosen standardized set of programmable settings is at least as effective as physician-tailored choices, as measured by the shock-related morbidity of implantable cardioverter-defibrillator (ICD) therapy.

BACKGROUND

Programming of ventricular tachyarrhythmia (ventricular tachycardia [VT] or ventricular fibrillation [VF]) detection and therapy for ICDs is complex, requires many choices by highly trained physicians, and directly influences the frequency of shocks and patient morbidity.

METHODS

A total of 900 ICD patients were randomly assigned to standardized (EMPIRIC, n = 445) or physician-tailored (TAILORED, n = 455) VT/VF programming and followed for 1 year.

RESULTS

The primary end point was met: the adjusted percentages of both VT/VF (22.3% vs. 28.7%) and supraventricular tachycardia or other non-VT/VF event episodes (11.9% vs. 26.1%) that resulted in a shock were non-inferior and lower in the EMPIRIC arm compared to the TAILORED arm. The time to first all-cause shock was non-inferior in the EMPIRIC arm (hazard ratio = 0.95, 90% confidence interval 0.74 to 1.23, non-inferiority p = 0.0016). The EMPIRIC trial had a significant reduction of patients with 5 or more shocks for all-cause (3.8% vs. 7.0%, p = 0.039) and true VT/VF (0.9% vs. 3.3%, p = 0.018). There were no significant differences in total mortality, syncope, emergency room visits, or unscheduled outpatient visits. Unscheduled hospitalizations occurred significantly less often (p = 0.001) in the EMPIRIC arm.

CONCLUSIONS

Standardized empiric ICD programming for VT/VF settings is at least as effective as patient-specific, physician-tailored programming, as measured by many clinical outcomes. Simplified and pre-specified ICD programming is possible without an increase in shock-related morbidity.

Для цитирования: Wilkoff Bruce L, Ousdigian Kevin T, Sterns Laurence D, Wang Zengri J, Wilson Ryan D, Morgan John M, EMPIRIC Trial Investigators. A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial.. Journal of the American College of Cardiology. 2006. DOI: 10.1016/j.jacc.2006.03.037.