MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/oto2.70269

Медицинская библиотека · OTO open · 2026

An Open-Source, Three-Dimensionally Printed, Motorized ("Breathing") Nasotracheoscopy Simulator.

Saad Anthony M, Herz David, Mehari Mal, Filimonov Andrey, Yan Kenneth, Kaye Rachel

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the face and content validity of a novel, 3D-printed nasotracheoscopy simulator with a motorized, moving larynx using expert feedback from board-certified otolaryngologists.

STUDY

DESIGN

Device development and validation study.

SETTING

An academic institution.

METHODS

An anatomically accurate airway model was created from a de-identified normal patient CT scan using open-source segmentation software. Negative impressions of the turbinates and nasopharynx were 3D printed in PLA and cast in silicone for realistic soft tissue. A custom-coded Arduino-controlled servo motor was integrated to simulate laryngeal motion. Ten board-certified otolaryngologists performed flexible nasotracheoscopy on the model and completed a survey assessing face validity, content validity, and general impressions using a 5-point Likert scale. Descriptive statistics summarized responses.

RESULTS

The simulator received high scores for face validity (mean 4.6, SD 0.67), content validity (mean 4.8, SD 0.40), and general impressions (mean 4.83, SD 0.42). All participants agreed that the model accurately represented anatomical landmarks and held educational value. Feedback emphasized its utility for telescopic navigation and anatomical recognition training.

CONCLUSIONS

This 3D-printed simulator demonstrated strong validity and realism. Its anatomical accuracy, dynamic features, and reproducibility support its use as a training adjunct in procedural training. Future studies will assess its effect on learner performance.

Для цитирования: Saad Anthony M, Herz David, Mehari Mal, Filimonov Andrey, Yan Kenneth, Kaye Rachel. An Open-Source, Three-Dimensionally Printed, Motorized ("Breathing") Nasotracheoscopy Simulator.. OTO open. 2026. DOI: 10.1002/oto2.70269.