DOI 10.1186/s12886-022-02402-1
Comparison of different pupil dilatation methods for phacoemulsification in eyes with a small pupil.
Аннотация
PURPOSE: To compare 6 methods for intraoperative pupil dilatation in eyes with insufficient pupil size during phacoemulsification.
METHODS
This was a prospective case-control study. 99 microcoria cataract patients (120 eyes) were collected and were divided into 6 groups(20 eyes each group), and their pupils were dilated by bimanual stretching pupil (group I), pupil radial cut open(group II), mechanical pupil dilatation with iris-retractor hooks (group III), OASIS iris expander (group IV), and Malyguin-ring (Microsurgical company, America) (group V), B-HEX Pupil Expander (Med Invent Devics, India)(group VI),respectively. 3.0 mm clear corneal incision were used in phacoemulsification. All cases were followed up at 1 week and 1, 3, 6 months after the surgery. The best corrected visual acuity (BCVA), intraocular pressure(IOP), corneal endothelium cell density(ECD), pupil diameter(PD) of before and after surgery were compared.
RESULTS
One same doctor finished all cataract surgeries successfully. The eyes' condition before surgery and at 6 months after surgery were compared. There were no significant statistical differences for the conditions of the eyes before surgery among six groups. The ECDs were better at 6 months postoperatively in group III and V, median values: 2114/mm2, 1961/mm2. PD was largest in group II (median value: 5.5 mm), which was significantly larger than other groups (Padjusted < 0.05).
CONCLUSIONS
All 6 methods used in this study were effective for the mechanical dilatation of small pupils and didn't affect the postoperative visual acuity and intraocular pressure in microcoria cataract phacoemulsification. Iris-retractor hooks and the Malyugin Ring can reduce intraoperative corneal endothelium cell loss. Postoperative PD is larger when the iris was cut open radially.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.