Report of a Cluster of Cases of Toxic Anterior-segment Syndrome After Implantation of a Specific Intraocular Lens Model.

Imamachi Katsue, Sugihara Kazunobu, Ikeda Yoshifumi, Matsuoka Yotaro, Tanito Masaki

Клинический случайСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

PURPOSE: To report a cluster of cases of toxic anterior-segment syndrome (TASS) in eyes implanted with a specific foldable acrylic intraocular lens (IOL) model.

DESIGN

Retrospective case series.

METHODS

The medical records were reviewed for 7 eyes of 4 patients diagnosed with TASS after cataract surgery at Shimane University and Matsue Red Cross Hospital between July and November 2020.

RESULTS

Among the 162 eyes implanted with the Lentis Comfort/LS-313 MF15 IOL, acute anterior chamber (AC) inflammation with fibrin formation developed 1-15 days after uneventful surgeries in 7 (4.3%) eyes (cataract surgery alone, n=4 eyes; combined cataract and minimally invasive glaucoma surgery, n=3 eyes). Other than local steroid use, fibrin membrane removal, YAG laser membranotomy, pars plana vitrectomy, and AC washout were performed to treat inflammation and/or secondary angle closure due to pupillary obstruction.

CONCLUSION

We experienced a cluster of TASS cases in eyes implanted with the Lentis Comfort/LS-313 MF15 IOL in a short period of time. To our knowledge, this is the first report of TASS associated with this IOL.

Для цитирования: Imamachi Katsue, Sugihara Kazunobu, Ikeda Yoshifumi, Matsuoka Yotaro, Tanito Masaki. Report of a Cluster of Cases of Toxic Anterior-segment Syndrome After Implantation of a Specific Intraocular Lens Model.. American journal of ophthalmology. 2021. DOI: 10.1016/j.ajo.2021.03.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.