A novel device for temporary cardiac stimulation using permanent pacemaker generators and active-fixation leads: the TEMPACE multicentre prospective study.
Аннотация
AIMS: This study aims to determine the safety and effectiveness of the KronoSafe® permanent pacemaker adapter for temporary cardiac pacing (TCP) with active-fixation leads (TCPAFLs).
METHODS AND
RESULTS
This was a multicentre, prospective, descriptive clinical investigation involving a medical device (ClinicalTrials.gov identifier: NCT05351658). Between January 2023 and December 2024, all consecutive patients who underwent implantation of a temporary active-fixation pacemaker were included. The devices were secured using the KronoSafe® adapter throughout the period of use. Complications during TCPAFL were recorded, and R-wave sensing, lead impedance, and pacing threshold were determined every 48 h. Thirty patients (10 with atrioventricular block, nine TAVI, five alcohol septal ablations, two slow atrial fibrillations, two bradycardia-tachycardia syndrome, and two pacemaker infections) who required TCP were included. Two centres participated in the study. The mean duration of TCP was 7.8 days (maximum 22 days), with 79.2% of the time spent in the cardiology ward. One complication (3%) was recorded, due to accidental traction of the system associated with an episode of agitation.
CONCLUSION
Temporary cardiac pacing with active-fixation leads using the KronoSafe® adapter is safe and effective, and its use can be extended outside the intensive care unit.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.