Медицинская библиотека · Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus · 2019

Activating the levator to elevate the eyelid.

Scott Alan B, Miller Joel M, Namgalies Talita Cunha

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To demonstrate in an animal model the feasibility of elevating the eyelid in a functionally useful manner by chronically stimulating the levator palpebrae superioris (LPS) muscle with an implanted electrode.

METHODS

Five rabbits were implanted with electrodes designed to stimulate the nerve innervating the LPS near its entry to the muscle. Bipolar platinum electrodes in a silicone rubber envelope with silicone-sleeved, PTFE-coated platinum lead wires were used to provide long-term stimulation with bipolar square-wave pulse trains of 0.18-0.80 mA and 200 Hz at a duty cycle of 8 seconds on and 2 seconds off. Explanted electrodes were examined for damage, and stimulated tissues were evaluated for abnormalities by light microscopy.

RESULTS

We achieved mean lid elevation of 1.6 mm, approaching the diameter of the light-adapted pupil, with 0.5 mA stimulus. Stimulus currents below 1.0 mA produced no signs of discomfort. Three animals with which we attempted daily stimulation, averaged 16.1 hours per week. Experiments lasted 22 weeks on average. Lid lifting with a well-implanted platinum electrodes was stable, with no apparent tissue or electrode damage after as long as 29.1 weeks.

CONCLUSIONS

Stable, functionally useful eyelid lifting was achieved with stimulation currents that caused no apparent discomfort or damage to muscles or nerves. A simple, discrete bipolar electrode was effective and survivable.

Для цитирования: Scott Alan B, Miller Joel M, Namgalies Talita Cunha. Activating the levator to elevate the eyelid.. Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus. 2019. DOI: 10.1016/j.jaapos.2019.04.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.