PURPOSE: To evaluate the post-approval long-term outcomes of the Argus II Retinal Prosthesis, with a specific focus on its functional visual benefit in patients' daily activities.

METHODS

Eighteen patients with bare light perception due to end-stage retinitis pigmentosa were included in a French prospective, multicentre, single-arm study and followed for 2 years. Visual benefit in patients' daily activities was monitored through the use of the Functional Low-vision Observer Rated Assessment (FLORA), and the final score at 2 years was the primary effectiveness outcome. Standardized visual assessments were also performed. Device- or procedure-related adverse events were recorded.

RESULTS

Seventeen subjects completed the study. Positive impacts of the Argus II system on functional vision and well-being were demonstrated for over 70% of subjects on the FLORA. Among the daily activities/tasks tested, finding doorways was one of the most statistically significantly improved tasks (p < 0.001), along with estimating the size of an obstacle (p < 0.001), visually locating a place setting on a dining table (p < 0.001) and visually locating people in a non-crowded setting (p < 0.001). Visual function was improved on most standardized tests. Only two device- or procedure-related serious adverse events were observed (one vitreous haemorrhage and one endophthalmitis, both resolved with treatment). No explantation was required.

CONCLUSION

This first report of a completed post-approval study of Argus II with a two-year follow-up demonstrates the safety and effectiveness of the Argus II System in a real-world cohort of patients and further highlights its real functional benefit in implanted patients' daily activities.

Для цитирования: Delyfer Marie-Noëlle, Gaucher David, Mohand-Saïd Saddek, Barale Pierre-Olivier, Rezaigua-Studer Fouzia, Ayello-Scheer Sarah, Dollfus Hélène, Dorn Jessy D, Korobelnik Jean-François, Sahel José-Alain. Improved performance and safety from Argus II retinal prosthesis post-approval study in France.. Acta ophthalmologica. 2020. DOI: 10.1111/aos.14728.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.