Smartwatch-Based Heart Rate Detection in Atrial, Ventricular and Pulseless Arrhythmias.

Turnbull Samual, Gupta Anunay, Garikapati Kartheek, Bhaskaran Ashwin, De Silva Kasun, Kotake Yasuhito, Bennett Richard G, Campbell Timothy, Thomas Stuart P, Chow Clara K, Kumar Saurabh

СтатьяПеревод в очереди

Аннотация

BACKGROUND & AIM: Smartwatches can provide a non-invasive heart rate (HR) monitoring to aid screening for atrial fibrillation, however, the accuracy of HR detections in the context of other arrhythmias requires further investigation. We aimed to assess the accuracy of HR measurements of two commonly used wearables at onset, during and termination of various induced and paced arrhythmias.

METHOD: Patients undergoing an electrophysiology study were fitted intraprocedurally with both the Apple Watch Series 4 (AW) and Fitbit Versa (FB). Detected HRs on the devices were recorded at initiation, during, and at termination of various paced and induced arrhythmias. True HR was confirmed via intracardiac electrograms and 12-lead electrocardiogram. Detected HRs on the AW and FB were assessed for accuracy, defined as HR with in ±10% of the true rate.

RESULTS

A total 322 cardiac rhythms from 32 patients were analysed. Both AW and FB demonstrated reasonable HR detection during sinus rhythm (100% vs 87%; p=0.04) and supraventricular tachycardia (SVT) (89% vs 86%; p=1). Both devices performed poorly at detecting SVT termination (62% vs 41%; p=0.04) and ventricular tachycardia (41% vs 16%; p=0.007). Neither device identified changes in HR during ventricular fibrillation or asystole.

CONCLUSIONS

Both the Apple Watch and Fitbit demonstrate reasonable accuracy in HR detection during regular rhythms of supraventricular origin, however, struggle with arrhythmia terminations and ventricular arrhythmias. Care should be taken when assessing HR trends for presence of cardiac arrhythmias.

Для цитирования: Turnbull Samual, Gupta Anunay, Garikapati Kartheek, Bhaskaran Ashwin, De Silva Kasun, Kotake Yasuhito, Bennett Richard G, Campbell Timothy, Thomas Stuart P, Chow Clara K, Kumar Saurabh. Smartwatch-Based Heart Rate Detection in Atrial, Ventricular and Pulseless Arrhythmias.. Heart, lung & circulation. 2026. DOI: 10.1016/j.hlc.2026.02.016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.