Energy-resolved attenuation behaviour on dual-source photon-counting CT: Implications for quantitative assessment of femoral bone marrow in multiple myeloma.

Heidemeier Anke, Scheiner Franzisca, Huflage Henner, Lennartz Simon, Kunz Andreas Steven, Feldle Philipp, Bley Thorsten Alexander, Grunz Jan-Peter

СтатьяПеревод в очереди

Аннотация

RATIONALE AND

OBJECTIVES

Quantitative CT assessment of multiple myeloma relies on threshold-based differentiation between physiological tissue and myeloma infiltration, yet bone marrow attenuation may be influenced by photon energy. The aim of this study was to evaluate the energy-dependent attenuation behaviour of femoral bone marrow on virtual monoenergetic images derived from dual-source photon-counting CT (PCD-CT) and its implications for quantitative analysis.

METHODS

This retrospective single-centre study included 103 patients who underwent unenhanced ultra-high-resolution whole-body PCD-CT for multiple myeloma assessment. VMI were reconstructed at 40-190 keV. Attenuation was measured in femoral bone marrow, adjacent muscle, and subcutaneous fat. Contrast-to-noise ratios were calculated and differences across energy levels were analysed using Friedman tests with Bonferroni correction.

RESULTS

Femoral bone marrow attenuation increased with higher keV from -83.4 HU at 40 keV to -58.7 HU at 190 keV, while fat attenuation increased more markedly and muscle attenuation decreased (all p < 0.001). As a result, the attenuation difference between marrow and fat declined substantially from 99.9 HU at 40 keV to 26.9 HU at 190 keV (p < 0.001). Although image noise decreased toward higher keV, CNR declined from 8.4 at 40 keV to 4.1 at 190 keV (p < 0.001).

CONCLUSIONS

Bone marrow attenuation on dual-source PCD-CT is energy-dependent and should not be interpreted independently of the selected monoenergetic level. Quantitative assessment requires contextualization relative to surrounding tissues rather than reliance on fixed attenuation thresholds.

Для цитирования: Heidemeier Anke, Scheiner Franzisca, Huflage Henner, Lennartz Simon, Kunz Andreas Steven, Feldle Philipp, Bley Thorsten Alexander, Grunz Jan-Peter. Energy-resolved attenuation behaviour on dual-source photon-counting CT: Implications for quantitative assessment of femoral bone marrow in multiple myeloma.. European journal of radiology. 2026. DOI: 10.1016/j.ejrad.2026.113001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.