DOI 10.3928/1081597x-20191010-03

Медицинская библиотека · Journal of refractive surgery (Thorofare, N.J. : 1995) · 2019

Corneal Viscous Properties Cannot Be Determined From Air-Puff Applanation.

Francis Mathew, Matalia Himanshu, Nuijts Rudy M M A, Haex Bart, Shetty Rohit, Sinha Roy Abhijit

СтатьяПеревод в очереди

Аннотация

PURPOSE: To assess whether corneal viscous properties are measureable with air-puff applanation in patients.

METHODS

The study had 312 normal eyes, 107 fellow eyes of patients with keratoconus, and 289 keratoconic eyes. The Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany) deformation data for all eyes were analyzed using two models. First, a standard linear solid model (SLM) assumed the cornea was an elastic material only. Second, a two-compartment Kelvin-Voigt model (KVM) assumed the cornea was a visco-elastic material. Corneal stiffness and viscosity were calculated. Further, the deflection amplitude was phase shifted virtually relative to the air-puff applanation force to assess whether the KVM was capable of detecting corneal viscous properties from air-puff applanation. This was similar in concept to measured viscoelastic deformations in other soft tissues. The hysteresis area was also calculated with deformation (cornea and whole globe) and deflection (cornea only) amplitude. The greater the magnitude of the hysteresis area, the greater was the magnitude of corneal viscosity (µc).

RESULTS

Both the SLM and KVM reported similar magnitudes of corneal stiffnesses (correlation coefficient > 0.99). However, for a given model, corneal stiffness was significantly different between normal, fellow, and keratoconic eyes (P = .001). From the KVM, the corneal viscosity was different between groups (P = .001) but was small in magnitude (order of 10-9). The deflection hysteresis area was also small in magnitude (order of 10-6). In contrast, the KVM detected significant corneal viscosity only when the deflection amplitude was virtually phase shifted with respect to the air-puff applanation force.

CONCLUSIONS

No significant corneal viscous response was detected in patients who had air-puff applanation. [J Refract Surg. 2019;35(11):730-736.].

Для цитирования: Francis Mathew, Matalia Himanshu, Nuijts Rudy M M A, Haex Bart, Shetty Rohit, Sinha Roy Abhijit. Corneal Viscous Properties Cannot Be Determined From Air-Puff Applanation.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2019. DOI: 10.3928/1081597x-20191010-03.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.